8
Dymax ACCU-CAL™ 50 Radiometer User Guide
Figure 5. Adapter Installation
4.
Attach the Lightguide Adapter to the Spot Lamp Lightguide by inserting the Lightguide into the
Lightguide Adapter until it bottoms out. Tighten the set screw when the Lightguide is installed.
5.
Press and release the On/Off Key on the Face Plate to turn the Radiometer on and off.
Figure 6. Attach Lightguide Adapter to Lightguide (Step 4)
Figure 7. Face Plate
6.
Press and release the Light Source Key to select the light source being measured. The different light
source options are:
Flood Lamp - For use when measuring UV intensity of a flood-lamp light source.
3-mm Lightguide - For use when measuring UV intensity at the end of a 3-mm lightguide.
5-mm Lightguide - For use when measuring UV intensity at the end of a 5-mm lightguide.
8-mm Lightguide - For use when measuring UV intensity at the end of a 8-mm lightguide.
7.
Press and release the Mode Key to select an operating mode. The different operating modes are:
Peak Intensity - Shows the highest UV intensity in mW/cm
2
seen by the detector during the
measurement.
Intensity - Shows the UV intensity in mW/cm
2
at the detector during the measurement.
Dose - Shows the total UV dose in mJ/cm
2
at the detector during the measurement.