Replacing Batteries (Battery Powered Model Only)

When the battery is low, a yellow warning LED will begin to flash and an
audible alarm will sound every 300 seconds to alert that it is time to
change the battery.  Using a large flat head screw driver, turn the battery
cover 1/4 turn in the direction of the arrows on the cover. Remove the
battery  cover  by  lifting  it  up.  Remove  the  coin  cell  battery  from  its
compartment.  Insert  a  new  3V  CR2450  coin  cell  battery  in  the
compartment with the positive (+) side of the battery facing up. Replace
the battery cover and twist the cover using the large flat head screw driver
in the opposite direction of the arrows on the cover.

Wiring

The wires in the 4.9 ft (1.5 m) cable for the Series WD3 are color coded
to reduce wiring errors. For battery powered models, Figure 2 has the
wiring diagram. For external powered models, Figure 3 has the wiring
diagram.

Operation

After mounting and wiring the Series WD3, it is ready for operation. The
water is detected when it bridges the two conducting strips on the bottom
of the unit. Upon detecting water, the relay contact will close and a red
LED  will  begin  to  flash.  On  models  with  the  built  in  annunciator,  the
annunciator  will  begin  to  chirp  until  the  water  bridge  between  the
conducting strips evaporates. After water evaporates, the relay contacts
will return to their normal state.

LED Indicators

For Model WD3-LP-D2 and WD3-LP-D2-A
Green – On when power is supplied to the unit
Red – On when water is detected 

For Model WD3-BP-D1-A
Yellow – On when low battery condition is detected
Red – On when water is detected

MAINTENANCE/REPAIR

Upon  final  installation  of  the  Series  WD3  no  routine  maintenance  is
required. The Series WD3 is not field serviceable with the exception of
replacing the batteries on battery powered model and should be returned
if repair is needed. Field repair should not be attempted and may void
warranty.

WARRANTY/RETURN

Refer  to  “Terms  and  Conditions  of  Sales”  in  our  catalog  and  on  our
website.  Contact  customer  service  to  receive  a  Return  Goods
Authorization  number  before  shipping  the  product  back  for  repair.  Be
sure  to  include  a  brief  description  of  the  problem  plus  any  additional
application notes.

DWYER INSTRUMENTS, INC. 

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. 

Fax: 219/872-9057

e-mail: [email protected]

©Copyright 2013 Dwyer Instruments, Inc.

Printed in China 7/13

FR# R3-444159-00 Rev.1

Model
WD3-BP-D1-A
WD3-LP-D2
WD3-LP-D2-A

Output

SPST NO SSR
DPDT Relay
DPDT Relay

Power

Battery
11 to 27 VAC/DC
11 to 27 VAC/DC

Audible Alarm

Yes
No
Yes

Figure 2: Wiring for WD3-BP-D1-A

Figure 3: Wiring for WD3-LP-D2 and WD3-LP-D2-A

YELLOW

GREEN

YELLOW

WHITE

GREEN

BLUE

BROWN

RED

BLACK

ORANGE

PWR IN