43
8.3
Use of surface-treated accessories
Eliminate pitting, corrosion, unevenness, etc. with paint thinner and sandpaper from the pipe
portion where the detector is to be mounted.
Note) In case jute is wound on a pipe, it should be peeled off before the above treatment.
When cast iron pipe is used, grind the sensor mounting surface by using a sander for
smoothness.
Jute is wound
Detector
Width
Pipe
Width
Small outer
diameter
PSX2-B
320 mm
(12.6")
or more
Small size
(standard) sensor
PSX2-A
540 mm
(21.3")
or more
Large size sensor
PSX2-D
Medium size sensor
PSX2-C
Mounting dimension
(L) + 200 mm
(7.9")
or more
Mounting dimension
(L) + 200 mm
(7.9")
or more
High temperature
PSX2-HT
530 mm
(20.9")
or more
Bulletin F-9-PUX2