OUTPUT CALIBRATION
1. Adjust the input frequency at least 15% below the maximum desired
input value, then adjust the input frequency for the desired MAXIMUM
output value.
2. Press the CAL switch. The red IN LED will turn off, and the yellow OUT
LED will stop flashing and stay on.
3. Adjust the input frequency at least 15% above the minimum desired
input value, and then adjust the input frequency for the desired MINIMUM
output value.
4. Press the CAL switch. The yellow OUT LED will turn off and the cali-
bration values will be stored in nonvolatile memory.
If the red and yellow LEDs flash after calibration, an error condition exits,
i.e. the input or output span was set too small, or the minimum/maximum
values were reversed.
If it is desired to retain the previous calibration values and exit calibration,
simply remove power to the unit before pressing the CAL switch at the
minimum input or output step.