PAIRING WIRELESS PROBES

1. Turn on Model UHH Universal Handheld by pressing the        button.
2. Press the      and      buttons to scroll through the menu headings at the top of the
display.
3. When 

PROBE

is highlighted, hit the        button to access the probe menu.

4.  Press  the          to  scroll  through  the  sub-menu  headings.  The  current  selected
parameter will be highlighted in yellow.
5. When 

PAIRING MODE

is highlighted, hit the        button to access the pairing mode.

6. Turn on the wireless probe(s) to be paired. After a period of up to 15 to 20 seconds,
the  UHH  screen  will  update  with  the  information  about  the  wireless  probe(s)  just
turned on.

7. Press the       button to scroll through the available probes. The current selected
probe will be highlighted in yellow.
8. When desired probe to be paired is highlighted, hit the       button to pair the probe.
Once it is paired, it will be removed from the list automatically.
9. Once all the desired probes are paired, press      button.
10. Repeat step 9 to go back to the home screen and begin readings.

SELECTING A PROBE

In  order  to  cycle  through  the  probes  that  are  paired  to  the  Model  UHH  Universal
Handheld, press the      or      button while in the view data mode.

SETTINGS

When using the Model VP2 Wireless 100 mm Thermo-Anemometer Probe, the base
unit can display velocity or air flow along with temperature and humidity. The settings
allow users to select velocity or air flow, engineering units, measurement range, and
which parameters to show on the display. To access the setting menus:

1. Press the      or      arrows to scroll through the menu headings at the top of the
display.
2. When 

PROBE

is highlighted, hit the        button to access the probe menu.

3. Press the       arrow to scroll through the sub-menu headings. The currently selected
parameter will be highlighted in yellow.

Flow / Velocity Selection

1. When sub-menu 

TYPE

(next to Anemometer) is highlighted, hit the       button to

access the settings for the sub-menu.
2.  Press  the        arrow  to  scroll  through  the  parameters.  The  currently  selected
parameter will be highlighted in yellow.
3. When 

DISPLAY

is highlighted, press the      button and velocity or vol. flow will be

highlighted.
4. Pressing the      or      buttons will alternate between velocity and vol. flow.
5. Once the desired selection is made, press      button.

Units Selection

1.  When  sub-menu 

TYPE

(next  to  Anemometer,  Humidity  or  Temperature)  is

highlighted, hit the        button to access the settings sub-menu.
2. Press the      arrow to scroll through the parameter headings. The currently selected
parameter will be highlighted in yellow.
3.  When 

UNITS

is  highlighted,  press  the          button  and  the  current  units  will  be

highlighted.
4. Pressing the       or      buttons will cycle through the available units.
5. Once the desired selection is made, press       button.

Area Adjustment (Only when Display is set to Volumetric Flow)

1. When sub-menu 

TYPE

(next to Anemometer) is highlighted, hit the      button to

access the settings sub-menu.
2. Press the      arrow to scroll through the parameter headings. The currently selected
parameter will be highlighted in yellow.
3. When 

AREA

is highlighted, press the       button to enter a new submenu that allows

selection of the shape of the duct, engineering units in which the duct is measured,
and the dimensions of the duct.
4. Press the      arrow to scroll through the parameter headings. The currently selected
parameter will be highlighted in yellow.
5. When the desired menu is highlighted, press the       button and the current value
of the parameter will be highlighted.
6.  Pressing  the        or        buttons  will  cycle  through  the  available  options  for  each
parameter.
7. Once the desired selection is made, press      button.
8. After all of the area parameters are entered, press      button.

Display Measurements

1.  When 

SOURCE

is  highlighted,  press  the        button  and  the  measurement  type

selected will be highlighted.
2. Pressing the      or       buttons will cycle through the available parameters.
3. Once the desired selection is made, press      button.
4.  In  order  to  turn  off  the  secondary  displays,  use  the        and      to  select  the
measurement type that is not selected for the 

SOURCE

5. When the desired 

TYPE

is highlighted, hit the        button to access the settings

sub-menu.
6. Press the      arrow to scroll through the parameter headings. The currently selected
parameter will be highlighted in yellow.
7. When 

VISIBLE

is highlighted, press the     button and either ON or OFF will be

highlighted. The measurement will be displayed when set to ON and not displayed
when set to OFF.
8. Pressing the       or      buttons will alternate between ON and OFF.
9. Once the desired selection is made, press      button. 

The  measurements  on  the 

MAJOR

display  will  always  be

displayed,  but  the  measurements  on  the  secondary  displays

can be turned off.

NOTICE

If a probe does not appear, power the probe down, then power
it back on.

NOTICE

Содержание IC-VP2

Страница 1: ... it will not pair while in low battery condition Step 1 Open the USB cover on the bottom of the probe handle Step 2 Plug the mini USB connector end of the cable into the probe handle Step 3 Plug the USB connector end of the cable into the port on the charger or PC Step 4 Plug the charger into either an electrical outlet or car charging port LED on the charger and the handle should both light up St...

Страница 2: ... sub menu TYPE next to Anemometer Humidity or Temperature is highlighted hit the button to access the settings sub menu 2 Press the arrow to scroll through the parameter headings The currently selected parameter will be highlighted in yellow 3 When UNITS is highlighted press the button and the current units will be highlighted 4 Pressing the or buttons will cycle through the available units 5 Once...

Страница 3: ...he and arrows to select the desired log file 6 Pressing the VIEW soft key will allow the user to view the first data point The units sensor time and date will be shown for each data point 7 Pressing the and arrows will cycle through all of the data points in the selected file 8 Data points can be deleted by pressing the DEL soft key 9 Pressing the BACK soft key will exit out of the file 10 Pressin...

Страница 4: ... 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail info dwyermail com Copyright 2014 Dwyer Instruments Inc Printed in U S A 5 14 FR 02 444116 00 Rev 3 Supplied in Australia by Instrument Choice Call our scientists on 1300 737 871 www instrumentchoice com au ...

Отзывы: