INSTALLATION
Mounting

A vertical position required. That is the position in which all standard models are
calibrated at the factory. Provide a 4-9/16˝ diameter opening in panel. Insert gage
and secure in place with provided screws and adapters.

Included Accessories:
Mounting lugs – 3
6-20 x 2-1/2 screws – 3
4-20 x 3/8 screws – 3 (mounting hole depth is 3/8˝ [9.53])
Cable gland – 1

Pressure Connections

Two integral tubing connections are provided on the back of the gage. They are
sized  to  fit  3/16˝  (4.8  mm)  I.D.  x  5/16˝  (7.9  mm)  O.D.  flexible  PVC  tubing.  To
measure single positive pressure, connect tubing to the + port and vent the – port
to atmosphere. To measure single negative pressure (vacuum), connect tubing to
the  – port  and  vent  the  +  port  to  atmosphere.  To  measure  differential  pressure,
connect higher pressure to the + port and lower pressure to the – port. Be sure the
pressure rating of the tubing exceeds that of the operating ranges.

Electrical Connections

A 9V battery or an external power supply 9-24 VDC can be used to power the unit.
If both battery and external power supply are connected, the battery will work as
back up power.

Battery Installation

The unit is shipped with a separate 9V alkaline battery. Remove the unit’s top back
cover, do not remove bottom cover. Connect the 9V battery to the battery holder as
shown in Figure 1. Replace top back cover and tighten screws to 2 ±.25 in. lbs.

External Power Supply Installation

Remove  the  unit’s  back  cover  then  connect  the  external  power  supply  to  the
terminal block as shown in Figure 1.

OPERATING INSTRUCTIONS
Key Functions

Home Position
Function

Allows access to the
menus

Displays pressure
readings instantly 
(rAtE KEY MODE only)

Displays full scale
range of unit

Main Menu
Function

Return to home
position

Sequences through
menus

Sequences through
menus

Enter into SUB
MENU

Sub Menu
Function

Return to previous
menu

Increments a value

Decrements a value

Changes a value or
setting. Press
ENTER and display
will blink. Adjust with
UP or DOWN
arrows. Press
ENTER to store.
Display will stop
blinking.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ø5/8 [15.88]

THRU HOLE FOR

ONDUIT CONNECTION

OUTSIDE DIAMETER

OF A-286 FLANGE

1-1/4 [31.75]

(3) FLANGE MOUNTING
HOLE LOCATIONS

Ø

 

B

 

 

 

 

 

 

FIGURE 1

Menu

Up Arrow

Down Arrow

Enter

POWER MUST BE OFF WHILE WIRING CONNECTIONS
ARE BEING MADE. DO NOT EXCEED SPECIFIED SUPPLY

VOLTAGE RATINGS. PERMANENT DAMAGE NOT COVERED BY WARRANTY
WILL RESULT.

CAUTION

www.

.com

[email protected]

1.800.561.8187

Содержание DigiMag DM-1112

Страница 1: ...st or interrupted SPECIFICATIONS Service Air and non combustible compatible gases Wetted Materials Consult factory Housing Materials Glass filled plastic Accuracy 1 F S including linearity hysteresis and repeatability 2 F S for ranges 1 w c and below Temperature Limits 0 to 140 F 18 to 60 C Compensated Temperature Limits 32 to 122 F 0 to 50 C Long Term Stability 1 F S per year Thermal Effect 0 05 ...

Страница 2: ... up power Battery Installation The unit is shipped with a separate 9V alkaline battery Remove the unit s top back cover do not remove bottom cover Connect the 9V battery to the battery holder as shown in Figure 1 Replace top back cover and tighten screws to 2 25 in lbs External Power Supply Installation Remove the unit s back cover then connect the external power supply to the terminal block as sh...

Страница 3: ...Menu Map www com information itm com 1 800 561 8187 ...

Страница 4: ...n SUB MENUS The Operation SUB MENUS for Series DM 1100 is PrES INWC Pressure Unit PRESSURE SUB MENU PRES INWC PrES INWC Pressure inches of water column With the LCD reading PrES INWC press the E key the display will blink Press the q key to change the pressure unit then press the E key to save the desired unit For pressure measurement the following units are available INWC Inches of water column M...

Страница 5: ...ssure reading NOTES 1 Depending on the Display Update setting the battery life is shown below 150 hours typical if Display Update is set for Normal 1 second update rAtE NORM 9 months typical if Display Update is set for 10 minutes rAtE 10 1 5 years typical if Display Update is disabled rAtE KEY 2 If the Display Update is set for 10 minutes or disabled On Touch mode the process pressure value can b...

Страница 6: ...STANDARD MOUNTING HOLE DIMENSION FOR DIGIMAG GAUGE Ø4 9 16 115 89 1 25 44 2X30 2xØ13 32 10 32 THRU HOLES FOR PRESSURE PORTS 1 51 64 45 72 3 15 32 88 12 3 SURFACE MOUNTING HOLE LOCATIONS 13 32 10 16 Ø5 8 15 88 THRU HOLE FOR CONDUIT CONNECTION OUTSIDE DIAMETER OF A 286 FLANGE 1 1 4 31 75 3 FLANGE MOUNTING HOLE LOCATIONS Ø5 17 32 140 49 BOLT CIRCLE MOUNTING HOLE DIMENSION FOR A 286 FLANGE Ø5 1 8 130 ...

Отзывы: