Dwyer Instruments DCG Series Скачать руководство пользователя страница 1

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. Box 373 • Michigan City, IN  46361 U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

The Dwyer Series DCG Digital Calibration Gage 

features a

remarkable 0.05% full-scale accuracy that includes the effects of

linearity, hysteresis, repeatability, and temperature across the

entire compensated temperature range. This gage was specifical-

ly designed to be a portable test gage where its unmatched accu-

racy is required. This is a highly durable gage that is housed in a

rugged and compact 3 inch stainless steel case.   

This highly accurate test gage offers menu configurable and pass-

word protected features that include engineering unit conversion,

seven selectable languages, field calibration capabilities, as well as

an adjustable bar graph and update/dampening rates.

INSTALLATION

When installing the gage use a 7/16˝ wrench on the wrench flat of

the gage to tighten the gage to the process.  DO NOT tighten the

connection by applying torque to the gage body which could

cause permanent damage and render the gage inoperable.  It is

good practice to apply Teflon

®

tape or pipe sealant to the gage

threads. 

Teflon

®

is a registered trademark of E.I. du Pont de Nemours and Company

Series DCG Digital Calibration Gage

Specifications - Installation and Operating Instructions

Bulletin  A-34-DCG

SPECIFICATIONS

Service:

Compatible, non-combustible liquids & gases.

Wetted Materials:

316 SS.

Housing Materials: 

300 Series SS.

Accuracy:

0.05% F.S. including linearity, hysteresis, repeatability,

and thermal effects across entire compensated temperature

range.

Pressure Limit: 

2 x F.S. range.

Temperature Limits: 

0 to 150°F (-18 to 66°C).

Compensated Temperature Range:

0 to 150°F (-18 to 66°C).

Size:

3.00˝ OD x 1.70˝ Deep.

Process Connection:

1/4˝ male NPT.

Enclosure Rating: 

NEMA 4 (IP65).

Display:

4-1/2 digit (0.66˝ high); 19999 counts.

Power Requirements: 

Three AAA batteries.

Battery Life: 

>1000 hours.

Auto Shut-off: 

Gage: On/off, 2 min., 5 min., 15 min., 30 min.;

Backlight: On/off, 10 sec., 30 sec., 1 min., 5 min.

Agency Approvals:

CE.

4-11/32

[110]

Ø3-11/32

[Ø85]

1-11/16

[43]

5/8

[15.5]

DCG iom  3/17/05  11:23 AM  Page 1

Содержание DCG Series

Страница 1: ...e to the process DO NOT tighten the connection by applying torque to the gage body which could cause permanent damage and render the gage inoperable It is good practice to apply Teflon tape or pipe sealant to the gage threads Teflon is a registered trademark of E I du Pont de Nemours and Company Series DCG Digital Calibration Gage Specifications Installation and Operating Instructions Bulletin A 3...

Страница 2: ...eturn the gage to measurement mode Note MIN MAX data is lost when unit is turned off After power up and before the pressure is applied this button will re zero the gage from any initial offset less than 5 of the gage s calibrated full scale range If indicated pressure is greater than 5 of the full scale pressure the re zero feature becomes inoperable This prevents accidental tare of a pressurized ...

Страница 3: ...e of the available time options Step 4 Press ENTER to finalize the selection LANGUAGE This menu option allows for selection of the lan guage to be used on the gage display English is the default set ting but the other options available include French Spanish German Italian Portuguese and Dutch To use the LANGUAGE option Step 1 Press the CONFIG key until the word LANGUAGE appears Step 2 Press ENTER...

Страница 4: ...ess ENTER Step 3 Enter the user defined password This is the same pass word established to access the CONFIG mode in the menu Step 4 Press the CONFIG key to select the desired option Step 5 Press ENTER to finalize the selection Note Selecting the DISABLD option does not disable the CLEAR key func tion for the MAX MIN feature If the DISABLD option is selected pressing the ZERO key will cause the di...

Отзывы: