Dwyer Instruments AQH-20 Скачать руководство пользователя страница 2

When desired value is reached  press the 

button to save the value and

return to the main menu or press 

button to return to main menu without

saving the changes

f the alarm sounds  simply press any key (but power key) to silence the alarm

Changing Engineering Units

Hold down 

for more than one second during normal mode to reach the main

menu

Press the 

or the 

to toggle between alarm setup and unit setup

When “Units” and “P3 0” are displayed on the LCD  press the 

button to

enter engineering unit setup

The current temperature unit will be blinking on the lower left of the display  Press
the 

or the 

to toggle between Celsius and Fahrenheit  

When desired unit is set  press the 

button to save and return to the main

menu or press  

button to return to the main menu without saving the changes

Holding the Current Value

Press and release the 

to freeze the primary temperature display  “HOLD”

will appear at the top left of the screen

n order to return to normal operation  press the  

button again and the “HOLD”

icon disappears from the display

Max, Min, and Average Data Functions

Press the  

button to cycle through the M N  MAX STEL and TWA functions

To exit   cycle  through  until  none  of  the  four  mentioned  functions  appear  on  the
screen

M N  and  MAX readings  show  the  minimum  and  maximum  CO

2

readings  on  the

main display  respectively  STEL and TWA readings show the weighted average of
the CO

2

readings on the main display  STEL uses all the CO

2

readings from the

past 15 minutes to calculate the average  TWA uses all the readings from the past
8 hours to calculate the average  The current temperature and humidity continues
to be displayed at the bottom of the LCD

Automatic Power Off (Sleep Mode)

The instrument will shut off automatically after 20 minutes of operation to conserve
battery life if no button is pressed  To disable the sleep mode  press “Ø set” and the
“HOLD”  buttons  simultaneously  before  power  on  An “n”  icon  will  appear  in  the
center of the display once disabled

BATTERY REPLACEMENT

The Air Quality Handheld displays 

as a visual low battery indication

When  “Lob”  appears  on  the  LCD  and  a  beeper  sounds   the  batteries  must  be
replaced

To Replace the Battery

1  Make sure the unit is powered off
2  Lay face down on a clean  flat surface
3  Open battery compartment by pushing in tab and lifting cover
4  Replace battery  taking note of the indicated polarity
5  Replace the cover

CO

2

Calibration

The CO

2

sensor should be calibrated in fresh outdoor air that is well ventilated and

in good weather conditions  Do not calibrate in areas with high CO

2

concentrations

such as crowded areas  

Procedure
1  Place meter in the calibration site
2  Turn the unit on and hold down 

and 

simultaneously to enter CO

2

calibration mode
3  “400ppm” and “CAL” should be blinking on the display  
4  Wait 10 minutes until the blinking stops and the unit returns to normal mode

To abort the calibration  turn the meter off at any time

RH Calibration

The RH must be calibrated with the default calibration salt   f not done properly and
with the correct salts it will cause permanent damage  

33% Calibration Procedure
1  Plug sensor probe into 33% salt bottle  
2  Hold down 

and 

under normal mode to enter 33% calibration  

3   “CAL”  and  the  calibrating  value  should  be  blinking  on  the  LCD  with current
temperature displayed but not blinking
4  Wait about 60 minutes until “CAL” and the calibrating value stop blinking

75% Calibration Procedure
1  After 33% calibration  plug the sensor probe into 75% salt bottle
2  Press the 

button to enter 75% calibration mode

3   “CAL”  and  the  calibrating  value  should  be  blinking  on  the  LCD  with current
temperature displayed but not blinking
4  Wait about 60 minutes until “CAL” and the calibrating value stop blinking and the
meter returns to normal mode  

Users can also calibrate either calibration point individually  For the 33% calibration
point   press  the 

once  the  33%  calibration  is  complete For  the  75%

calibration point  press the 

or 

within the first five minutes while

initializing the 33% calibration

MAINTENANCE

A periodic check of the system calibration is recommended  Model AQH-20 is not
field serviceable and should be returned if repair is needed (field repair should not
be attempted and may void warranty)  Be sure to include a brief description of the
problem plus any relevant application notes  Contact customer service to receive a
return goods authorization number before shipping

©Copyright 2013 Dwyer  nstruments nc

Printed in U S A  8/13

FR# R6-443744-00 Rev 2

M  AV

N

X

M /AV

N

X

M  AV

N

X

CAL

Esc

CAL

Esc

CAL

Esc

DP/WBT

MODE

DP/WBT

M /AV

N

X

CAL

Esc

MODE

HOLD

CAL

Esc

HOLD

The RH calibration should be done at 25ºC and humidity close
to the calibration salt being used (must at least be stable)

NOTICE

Model AQH-20  requires  at  least  5  minutes  worth of  values  in
order to calculate either of the CO

2

averages

NOTICE

f the unit fails to power on  it is likely that the batteries also need

to be replaced

NOTICE

Failing to turn the handheld off may result in malfunction of the
unit   Dispose  of  used  batteries  promptly  and  keep  away  from

children

NOTICE

M  AV

N

X