Dwyer Instruments 477A-1 Скачать руководство пользователя страница 2

INSTRUCTIONS

Battery Installation

The unit is shipped with a separate 9 volt alkaline battery which must be installed before

operation. Remove the two screws holding the bottom endcap in  place and remove the
endcap. Connect the battery to the enclosed battery clip observing correct polarity. Be
careful not to trap wires between the battery, case or foam pads which retain the battery.
This could make it difficult to install the battery or remove it later for replacement. Be sure
the rubber gasket is properly seated in the gasket channel of the endcap and replace end-
cap. Note that the endcap will only fit one way because the holes are slightly off-center.
Place the “Z” shaped wrist strap clip in one of the screw recesses and replace the screws.
Do not overtighten the screws. Attach wrist strap to clip.

When battery replacement becomes necessary, use only a 9 volt alkaline type such as

a Duracell

® 

MN1604, Eveready

®

522 or equivalent. Zinc-carbon types, often labeled Heavy-

duty are not recommended because of the increased potential for leakage. Alkaline batter-
ies are also a better value because they last up to three times longer in this device.

On-Off Operation

The on-off control is a toggle function. Press and release the ON/OFF key once to turn

unit on; again to turn it off. If the manometer is left on with no activity for approximately 20
minutes, unit will turn itself off to conserve the battery.

Display Backlight

The Model 477A includes a display backlight to allow use in the dark or in poor lighting con-

ditions. Manometer must be switched off before this feature can be activated. Next, press and
hold the ON/OFF key down. After about 1 second the backlight will come on and remain light-
ed for approximately 2 minutes after which it will turn itself off to conserve battery life.

Zeroing Pressure Reading

Potential inaccuracy due to temperature effects can be minimized by re-zeroing immediately

before use. To zero the display, vent both ports to atmosphere so no pressure is applied to
either port. Press the ZERO/STORE key and - - - - will be momentarily displayed as zeroing
occurs. Zeroing is not possible when the memory mode is in use. It must be done before select-
ing that function.

If the unit is accidently zeroed with pressure applied to one of the ports, the pressure read-

ing might display incorrectly. To correct, vent the pressure ports to atmosphere and press the
ZERO/STORE key to zero the unit.

Pressure Connections

To measure single positive pressure, connect tubing to port and vent oppo-

site port to atmosphere. To measure differential positive pressure, connect higher positive
pressure to port and lower positive pressure to port marked -. Manometer will
indicate the difference between the two.

Selecting Pressure Units

Up to seven pressure units are available. The display will indicate the current selection.

To change to different units, use the UNITS/LOC key. Each touch will cause an advance to
the next choice. The selected units will remain in memory even when power is shut off. This
way, your preference will always be displayed after the initial selection.

D-26:D-26(478)  7/10/09  10:04 AM  Page 2

Содержание 477A-1

Страница 1: ...ax 219 872 9057 e mail info dwyer inst com SPECIFICATIONS Service Air and compatible gases Wetted Parts Consult factory Accuracy 0 10 of full scale from 60 to 78 F 15 6 to 25 6 C 1 of full scale from 32 to 60 and 78 to 104 F 0 to 15 6 and 25 6 to 40 C Pressure Hysteresis 0 1 of full scale Pressure Limits See chart Temperature Limits 32 to 104 F 0 to 40 C Storage Temperature Limits 4 to 176 F 20 to...

Страница 2: ...es a display backlight to allow use in the dark or in poor lighting con ditions Manometer must be switched off before this feature can be activated Next press and hold the ON OFF key down After about 1 second the backlight will come on and remain light ed for approximately 2 minutes after which it will turn itself off to conserve battery life Zeroing Pressure Reading Potential inaccuracy due to te...

Страница 3: ...reading press and hold the HOLD MEMORY until ST01 is displayed then release the key Next press ZERO STORE key to save current reading to ST01 memory location A beep will sound indicating that the reading has been saved As each reading is saved the mem ory location display will advance to the next number To resume pressure measurement press the HOLD MEMORY key again Note that in the memory mode the...

Страница 4: ... unit Exceeding the range will not damage it or affect calibration as long as the maximum rated pressure is not exceeded Do not exceed the maximum rated pressure of the manometer Doing so will cause permanent damage to the sensor may rupture the housing and or cause injury The maximum pressure is shown on the rear label and on page 1 of these instructions Low Battery Indicator A weak battery can c...

Отзывы: