CALIBRATION

IMPORTANT NOTE:

Electrochemical CO sensors slowly lose sensitivity over

time. A quick “bump test” should be performed with CO gas before each usage

to ensure proper instrument performance.  It is also recommended to perform

a calibration at four month intervals to guarantee proper operation. A replace-

ment sensor should be installed if the unit can no longer be calibrated. In order

to perform a calibration, a calibration kit is required. This is available from Dwyer

Instruments, see below to order.

CALIBRATION ACCESSORIES

A-450-1 Calibration Kit with Regulator:

Includes case, Series 450 Handheld

Carbon Monoxide Monitor, regulator, tubing and A-450-2 Calibration Adapter.

For use with 17 liter gas tank.  The regulator has a fixed flow rate of 0.5 L/min.

If a different Calibration Kit is used, a flow meter is required that can verify set-

ting the flow rate correctly to 0.5 L/min.

A-450-2 Calibration Adapter:

Adapter connects the regulator to the sensor

cover on the back of the CO Monitor.

A-450-3 Replacement Sensor

A-450-4 Aspirator Kit: 

Includes aspirator bulb, adapter, tubing and probe.

A-450-5 Calibration Regulator Kit

: without Series 450 Carbon Monoxide

Monitor (same as A-450-1 except without Series 450 Handheld Carbon

Monoxide Monitor.

CALIBRATION PROCEDURE

1. Zeroing the Instrument: It is important that the instrument is in an area with

zero concentration of CO gas. Turn the instrument on and allow it to stabilize for

three (3) minutes. Press the         key until the display reads CAL0. The upper

display should read between 0 and 12. If the upper display reads above 12, the

sensor may be bad and should be replaced – refer to SENSOR REPLACE-

MENT. Press the 

and          keys simultaneously. The display will momen-

tarily blink to show the value has been stored.

5

HOLD

STORE

ENTER

ON

OFF

BACKLIGHT

INDEX

HOLD

STORE

ENTER

ON

OFF

BACKLIGHT

INDEX

HOLD

STORE

ENTER

ON

OFF

BACKLIGHT

INDEX

d-25small  11/5/02  3:02 PM  Page 5

Содержание 450 Series

Страница 1: ...n labeled Heavy Duty are not recommended because of their shorter life and increased potential for leakage Alkaline batteries are a better value because they typically last up to three times longer in this device Duracell is a registered trademark of Duracell USA a division of Duracell Inc Eveready is a registered trademark of Eveready Battery Company Inc The Series 450 CO Monitor measures car bon...

Страница 2: ...nnunciator and the display will again be updated To Store a Reading In Memory Up to 25 CO readings may be stored in memory Pressing will store the presently displayed value in memory The active memory location will be displayed in the lower left starting initially with 1 The MEM descriptor will turn on any time there are any values stored in memory m MEM PPM CO Memory Location HOLD PPM CO HOLD Ann...

Страница 3: ...y shows the active memory location upper display shows the memory value Use and to sequence through the memory locations To clear the memory press and hold the key for five seconds while in this menu loca tion Reading the Peak Value The Peak feature stores the highest reading the monitor has measured since the last reset or Power On At Power On PEA is reset to the present reading To view the peak ...

Страница 4: ...LF Alarm Flasher is displayed in the lower left display Then use the key to turn the Visual alarm off The upper display will read OFF Then press to store the change To turn the Visual alarm back on press the key then press to store the change The Audio alarm may be turned off or on in the same way as the Visual alarm by indexing the menu until the lower left display reads ALb Alarm Beeper Zero Adj...

Страница 5: ...low rate correctly to 0 5 L min A 450 2 Calibration Adapter Adapter connects the regulator to the sensor cover on the back of the CO Monitor A 450 3 Replacement Sensor A 450 4 Aspirator Kit Includes aspirator bulb adapter tubing and probe A 450 5 Calibration Regulator Kit without Series 450 Carbon Monoxide Monitor same as A 450 1 except without Series 450 Handheld Carbon Monoxide Monitor CALIBRATI...

Страница 6: ...il the display reads CAL0 Press and hold the and enter keys for five seconds until the lower dis play reads CAL1 Verify that the upper display reads between 125 and 400 then press the and keys simultaneously The display will blink to show the calibration value has been stored If the value is 125 or greater than 400 the sensor is bad needs to be replaced Refer to SENSOR REPLACE MENT 4 Index the men...

Страница 7: ...e shorting wire spring from pins on new sensor 5 Carefully install new sensor into instrument Re install cap and mounting screws 6 New sensors must be recalibrated Wait a minimum of 30 minutes after installing a new sensor before calibrating the instrument The instrument has internal temperature compensation and the sensor must be at the same tem perature as the instrument to calibrate properly GA...

Страница 8: ...__ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________...

Отзывы: