Dwyer Instruments 450 Series Скачать руководство пользователя страница 1

Series 450 Carbon Monoxide Monitor

Instructions and Operation

SPECIFICATIONS
Range:

0-2000 PPM.

Resolution:

1 PPM.

Accuracy:

(using 200 PPM calibration gas)

±3% of reading ± the accuracy of the cali-
bration gas, at the time of calibration.

Response Time:

<30 seconds to 90% of

reading.

Operating Temperature:

32° to 104°F 

( 0° to 40°C).

Humidity Conditions:

0-90% Relative

Humidity Non Condensing. 

Adjustments:

Zero and Span via keypad.

Power*:

9V Battery.

Battery Life:

500 hours typical

Case Material:

Anodized Aluminum with

plastic end caps.

Dimensions:

6.8˝ L x 2.8˝ W x  0.9˝ D

except add 0.5˝ D for sensor cover.  
(172.7 x 71.1 x 22.9 mm) (12.7mm).

Sensor Type:

Field replaceable

Electrochemical Cell.

Sensor Life:

2 years minimum.

Alarms:

2 Alarms, with audible and visual.

DWYER INSTRUMENTS, INC. 

Phone: 219/879-8000     www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.    

Fax: 219/872-9057          e-mail: [email protected]

Lit-By Fax: 888/891-4963 

replace the endcap. Note that the endcap will
only fit one way because the holes are slightly
off-center. Place the “Z” shaped wrist strap clip
in one of the screw recesses and replace the
screw. Do not overtighten. Attach the wrist
strap to the clip.

Use only 9 Volt alkaline type batteries such as
Duracell

®

MN1604, or Eveready

®

522, or

equivalent. Zinc-carbon types, often labeled
Heavy-Duty are not recommended because of
their shorter life and increased potential for
leakage. Alkaline batteries are a better value
because they typically last up to three times
longer in this device.

Duracell

®

is a registered trademark of Duracell USA, a division of

Duracell Inc.
Eveready

® 

is a registered trademark of Eveready Battery Company Inc.

The Series 450 CO Monitor

measures car-

bon monoxide from levels 0 to 2000 PPM. The
unit is equipped with audible and visual indica-
tors of CO concentration. The beacon indica-
tor flashes in correspondence to the level of
CO detected in the air. Perfect for HVAC appli-
cations such as servicing heaters, furnaces,
appliances and for general building inspection.

BATTERY INSTALLATION

Remove two screws holding the bottom end-
cap labeled BATTERY in place and remove it.
Connect the battery to the enclosed battery
clip observing correct polarity. Insert the bat-
tery in the case. Be careful not to trap the
wires between the case or foam pad, which
retains the battery. This could make it difficult
to install the battery or remove it later for
replacement. Be sure the rubber gasket is
properly seated in the gasket channel and

Bulletin D-25

*NOTE:

This instrument is shipped with a new 9V bat-

tery. However, there is a protective plastic cap installed on
the battery terminals. This cap must be removed and the
battery connected to the battery terminal before the instru-
ment will function.

d-25small  11/5/02  3:02 PM  Page 1

Содержание 450 Series

Страница 1: ...n labeled Heavy Duty are not recommended because of their shorter life and increased potential for leakage Alkaline batteries are a better value because they typically last up to three times longer in this device Duracell is a registered trademark of Duracell USA a division of Duracell Inc Eveready is a registered trademark of Eveready Battery Company Inc The Series 450 CO Monitor measures car bon...

Страница 2: ...nnunciator and the display will again be updated To Store a Reading In Memory Up to 25 CO readings may be stored in memory Pressing will store the presently displayed value in memory The active memory location will be displayed in the lower left starting initially with 1 The MEM descriptor will turn on any time there are any values stored in memory m MEM PPM CO Memory Location HOLD PPM CO HOLD Ann...

Страница 3: ...y shows the active memory location upper display shows the memory value Use and to sequence through the memory locations To clear the memory press and hold the key for five seconds while in this menu loca tion Reading the Peak Value The Peak feature stores the highest reading the monitor has measured since the last reset or Power On At Power On PEA is reset to the present reading To view the peak ...

Страница 4: ...LF Alarm Flasher is displayed in the lower left display Then use the key to turn the Visual alarm off The upper display will read OFF Then press to store the change To turn the Visual alarm back on press the key then press to store the change The Audio alarm may be turned off or on in the same way as the Visual alarm by indexing the menu until the lower left display reads ALb Alarm Beeper Zero Adj...

Страница 5: ...low rate correctly to 0 5 L min A 450 2 Calibration Adapter Adapter connects the regulator to the sensor cover on the back of the CO Monitor A 450 3 Replacement Sensor A 450 4 Aspirator Kit Includes aspirator bulb adapter tubing and probe A 450 5 Calibration Regulator Kit without Series 450 Carbon Monoxide Monitor same as A 450 1 except without Series 450 Handheld Carbon Monoxide Monitor CALIBRATI...

Страница 6: ...il the display reads CAL0 Press and hold the and enter keys for five seconds until the lower dis play reads CAL1 Verify that the upper display reads between 125 and 400 then press the and keys simultaneously The display will blink to show the calibration value has been stored If the value is 125 or greater than 400 the sensor is bad needs to be replaced Refer to SENSOR REPLACE MENT 4 Index the men...

Страница 7: ...e shorting wire spring from pins on new sensor 5 Carefully install new sensor into instrument Re install cap and mounting screws 6 New sensors must be recalibrated Wait a minimum of 30 minutes after installing a new sensor before calibrating the instrument The instrument has internal temperature compensation and the sensor must be at the same tem perature as the instrument to calibrate properly GA...

Страница 8: ...__ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________________________ _____________________________...

Отзывы: