Precision Pressure Measurement

The Dwyer Microtector

®

combines the time-

proven principles of the Hook Gage type
manometer and modern solid-state integrat-
ed circuit electronics.  It provides an inexpen-
sive means of achieving accuracy and
repeatability within ± .00025 inches water col-
umn throughout its 0 to 2 inches w.c. range.
It is truly a new standard in precision measur-
ing devices.

Principles of Operation

A pressure to be measured is applied to the
manometer fluid which is displaced in each
leg of the manometer by an amount equal to
1/2 the applied pressure.  A micrometer
mounted point is then lowered until it contacts
the manometer gage fluid. The instant of con-
tact is detected by completion of a low-power
A.C. circuit.  Current for this circuit is supplied
by a 1.5 volt penlight cell feeding two semi-
conductor amplifiers which act as a free-run-
ning multivibrator operating at a frequency of
approximately two kilohertz.  Completion of
the A.C. circuit activates a bridge rectifier
which provides the signal for indication on a
sensitive (0 to 50 microamps) D.C. microam-
meter.

2

LOW

PRESSURE

FLUID

LEVEL

OFF

ON

On indication of contact, the operator stops
lowering the point and reads the micrometer
which indicates one half the applied pressure.
By interpolating eight divisions (each being
.000125˝ w.c.) between .001 micrometer
graduations, a total accuracy of .00025 can
easily be achieved. The micrometer complies
with Federal Specification GGG-C-105A and
is traceable to a master at the NIST.

Locating and Opening

Stand the Microtector

®

and case on a firm flat

level surface.  Remove cover by releasing the
latches and lifting it straight up.  If it is neces-
sary to move the gage without case, handle
only the base plate or clear acrylic block.
(

CAUTION:

Do not handle gage by grasping

meter-electronic package housing Item 7 on
drawing.)

Microtector

®

Gage

D-57  10/5/05  3:59 PM  Page 2

Содержание 1430

Страница 1: ...red in accor dance with ASME B89 1 13 2001 and are traceable to a standard at the National Institute of Standards and Technology Three point mounting dual leveling adjustment and circular level vial assure rapid setup Durablock precision machined acrylic plastic gage body Sensitive 0 50 microamp D C meter acts as a detector and also indicates battery and probe condition Heavy 2 thick steel base pl...

Страница 2: ...lifiers which act as a free run ning multivibrator operating at a frequency of approximately two kilohertz Completion of the A C circuit activates a bridge rectifier which provides the signal for indication on a sensitive 0 to 50 microamps D C microam meter 2 LOW PRESSURE FLUID LEVEL OFF ON On indication of contact the operator stops lowering the point and reads the micrometer which indicates one ...

Страница 3: ...ing contact turn the point out of the fluid by turning the micrometer barrel counter clockwise to a reading of 010 or more Again watch the meter and this time lower the point by turning the micrometer barrel The point position where the meter pointer begins to move up scale is the zero position This position should correspond to the zero reading on the micrometer Adjust the point in relation to th...

Страница 4: ...and dirt with a clean rag reassemble and recheck meter operation 2 If the meter operation continues to be sluggish replace the size AA 1 5 volt battery Replace the battery at least once a year to avoid deterioration of battery and damage to gage Leakproof alkaline battery is recom mended To replace the battery remove center screw Item 10 located in the back of the electronic enclosure Cover Item 9...

Отзывы: