11
RMO500 Getting Started
Note:
In case of PC and RMOWin software use during the test, all performed
measurements will be registered. Results will be shown in a table in
Excel format with option of additional edit and graphic presentation.
Using RMOWin after testing for transfer of results from selected internal memory
location to PC, it is possible to transfer only the result of the last test.
To stop the test prematurely, press the
STOP
button at any time during the test. The
current will immediately drop to 0A.
6.4 Viewing the Test Results
Once a Single
test is finished, RMO500 automatically changes to the
Result
menu
to display the test results.
For Continuous Test only the last performed test is shown.
Note:
If the defined test current could not be reached, this could be due to a
too high contact resistance of the test object and/or of the current cables.
the test current value displayed here is the maximum test current reached.
Now there are two choices:
1.
Starting a new test with a different current value:
– press
the
ESC
button to change to the
Sing
menu,
–
set a new current value and/or another memory position,
–
confirm your entries by pressing the
ENTER
button,
–
start the test by pressing the
Ω
button.
2.
Starting a new test with the same current value:
– press
the
ESC
button to change to the
Sing
menu,
– press
the
ENTER
button to change to the
Test
menu,
– press
the
Ω
button to run a test.
The current value remains the same as in the previous
test, the memory position increases by one.
Figure 6-8:
The
Result
menu
showing the test
results
Max. test current reached
Memory
position
that holds
the results
Measured resistance of test object