17
RMO10T Getting Started
6.5 Printing the Test Results
RMO10T stores up to 100 test results. They can be printed using the
menu at memory positions
0 ... 99.
Go to
menu using
LEFT
and then
UP
button. Selecting first and last memory position and press
ENTER
button. The results that are located in the range between these memory locations, including the
chosen memory locations, will be printed.
Note:
It is not possible to print test results using the built-in printer from RMOWin-T4Software.
If printing is attempted the device will be blocked and a reset of the device is necessary.
Look of a RMO10T Test Report are given at figures below:
6.6 Duty Cycles
During tests, RMO10T generates a high DC current that heats up the test set. To prevent RMO10T from
overheating, certain duty cycles apply depending on the test current used.
Cooling time between two measurements is proportional to test duration time
.
A built-in control prevents
skipping these cooling times. If you try to start a test within the cooling period, the display shows "Wait"
and a timer counting down. After the cooling period has passed, start a new test using the set test
parameters by pressing the
OMEGA
button.
C u r r e n t T e s t
Test current
(A)
Maximal test
duration time (min)
Cooling time for
maximal test
duration time (min)
5mA, 10mA,
50mA,100mA, 500mA,
1A, 2A, 5A, 10A
60 0
Figure 6-22:
The
menu showing
first and last memory
position and Date and
Time of printing
Table 6-23:
Cooling time
between current
tests