8.2
Using the hygienic protective
cover
WARNING
Danger of cross contamination when
not using the hygienic protective
cover or when using the hygienic pro-
tective cover more than once
❯
Do not use the unit without the
hygienic protective cover.
❯
Do not use the hygienic protective
cover more than once (disposable
item).
Wear protective gloves when applying the
disposable hygienic protective cover.
❯
Hold the camera so that the optical element
faces down.
❯
Lift the white edge of the disposable hygienic
protective cover and slide the camera head
into the cover. The transparent plastic side
must face upwards.
❯
Stretch the disposable hygienic protective
cover an extra 2 - 3 mm so that the cover
presses tightly against the optical element.
❯
Carefully press the disposable hygienic protec-
tive cover against the optical window using
your fingers. Ensure that there are no air bub-
bles between the optical window and the dis-
posable hygienic protective cover.
❯
Hold the disposable hygienic protective cover
firmly on the white edge and pull off the trans-
parent plastic side in the direction of the cam-
era head.
❯
Pull off the paper underside from the camera
head in the direction of the handpiece.
8.3
Place on the spacer
The spacer is required for imaging with the Proof
and Proxi interchangeable heads.
WARNING
Danger of cross-contamination when
used without reprocessing or follow-
ing incorrect reprocessing
❯
Sterilise the spacer in the steam ster-
iliser (see "9 Reprocessing of the
device") before each use.
❯
Place the spacer onto the interchangeable
head from above. Ensure that the spacer does
not cover the optical element of the inter-
changeable head.
Usage
2109100026L02 1907V001
23