6.5
Information about connecting
cables
The diameter of the connections depends on the
current consumption, length of line and the ambi-
ent temperature of the unit. Information concern-
ing the current consumption can be found in the
Technical Data supplied with the particular unit to
be connected.
The following table lists the minimum diameters
of the connections in relation to the current con-
sumption:
Current consumption of
unit [A]
Cross-section
[mm
2
]
> 10 and < 16
1.5
> 16 and < 25
2.5
> 25 and < 32
4
> 32 and < 40
6
> 40 and < 50
10
> 50 and < 63
16
Control cable
24 V protective low voltage for:
– Hose manifold
– Place selection valve
– Spittoon valve
Installation type
Line layout (minimum
requirements)
Fixed installation
– Shielded sheathed cable
(e.g. (N)YM (St)-J)
Flexible
– PVC data cable with
shielded cable sheath-
ing, as used for tele-
communications and IT
processing systems
(e.g. type LiYCY)
or
– Lightweight PVC control
cable with shielded
cable sheathing
7
Installation
7.1
Combining devices safely
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e.g. due to leakage currents).
❯
Only connect units when there can be no
question of danger to operator or to patient.
❯
Only connect units when it is safe to do so and
when there is no risk of damage or harm to the
surroundings.
❯
If it is not 100% clear from the unit data sheet
that such connections can be safely made or if
you are in any doubt, always get a suitably
qualified person (e.g. the manufacturer) to ver-
ify that the setup is safe.
7.2
Setting up and connecting
the lower part of the cabinet
❯
Set up the lower part of the cabinet.
❯
Connect the exhaust air hose and suction
hose.
❯
Remove the transport protection.
❯
Attach the cover at the rear.
Assembly
14
A949100052L02 2104V003
EN
Содержание A949444000
Страница 1: ...EN Power Tower View Installation and operating instructions A949100052L02 A949100052L02 2104V003 ...
Страница 2: ......
Страница 29: ...Appendix 13 Connection media plan PC board mit with CA 4 Appendix A949100052L02 2104V003 27 EN ...
Страница 33: ......
Страница 34: ......
Страница 35: ......