COD.988250035_REV.0.0_ED.06/2020
47
EN
“WARRANTY ACTIVATION” SCREEN
on the website www.DÜRR DENTAL.it:
The screen will appear every 24 hours for 30 days. Please register your warranty on
DÜRR DENTAL’s website – www.duerrdental.com – within 30 days.When you press the
“A” button you will see the “Registration Conirmation” screen.
” sul sito
A
3.1 FIRST POWER-UP
- Press the illuminated main switch (5 ig. 1).
1 - Wait for the main screen to load
2 - Select the desired language.
3 - Warranty Activation” screen.
4 - Home page loading following the choices made during warranty activation.
Select the “07 open door” icon to open the door and extract the supplied accessories from the chamber.
EXAMPLE OF
LANGUAGES SCREEN
HOME PAGE SCREEN
01 “Sterilisation cycles” menu
02 “Test” menu
03 “Settings” menu
04 “Data Output” menu
05 Last sterilisation cycle run
06 Last test run
07 Unlock door
08 Water illing
Check that the accessories include:
- Documentation (instruction manual)
- The tray support
- 3 trays (Basic version)
- 1 tray extraction pincer
- 1 illing hose
- 1 drain hose
- 1 funnel
WARNING
- Insert the tray holder as shown in the igure
- The tray holder must be inserted into the chamber with foot “A” facing down and positioned at the bottom of the chamber. The
position of foot “A” must be respected both when using 3 trays and 5 trays
- The tray holder must always have rubber protection “B” mounted on the eight supports
If the autoclave is not filled with water from external tanks, DO NOT CONNECT it directly to the water mains, but connect it to a
demineralizer or water treatment system that will ensure the required water specifications (see 7.1). The water treatment system
must prevent water backflow to the mains.
Validate the installation in accordance with the following table:
– Vacuum test
– Autoclave automatic test
– Autoclave instrumentation calibration check
– Overtemperature trip test
– Thermometric test for small loads (*)
– Drying test for small loads (*)
– Steam penetration test for porous loads (*)
– Thermometric test for solid loads (*)
– Drying test for solid loads (*)
– Steam penetration test for solid loads (*)
– Thermometric test for a load chosen by the user
– Microbiological test for a load chosen by the user
(*) To be carried out with cycles of 121°C and 134°C