Check that there are no masses (no load)

calibration begins with a zero balance. 

When ready, press 

ENTER.

While the screen is displaying “Reading...”, 

the 9840 is calibrating. Wait for just a few 

With that information from the calibration 

buttons  

+

 , 

 , 

 , and 

.

Then press 

ENTER

.

For the next display, you will be 

required to enter the mV/V for the 

full-scale output of the direction that 

you are going to use. In this example, 

we will pick “Tension”.

We see that the value of 4.07906 is the 

full-scale mV/V, correlating with the max 

capacity of 10K that we entered earlier.

7418 East Helm Drive • Scottsdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com

Setup Guide

 9840  v

1.1

  0

9

-

18

-2018

 Page 

6

 of 

7

Содержание 9840

Страница 1: ... using In starting out you will see User Data Entry on the screen To enter setup mode press the two end buttons at the same time and 1 2 3 As it starts up you will see a series of screens introducing you to the 9840 Load Cell Connector and displays of the settings from its last use 4 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Page 1 of 7 Setup Guide 9840 v1 1...

Страница 2: ...ER After being brought to the next screen press till you get to the Calibration option and then press ENTER Before you can enter the calibration settings you will be required to enter a password 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 2 of 7 ...

Страница 3: ... the measurement cell type that you are using For this example we are using a load cell so we select Load Then press ENTER 10 Enter the cell s serial number by using the or signs then for the next digit ENTER 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 3 of 7 ...

Страница 4: ...git s Do the same for the day and year and press ENTER after each one Select the necessary unit of measure and press ENTER 11 12 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 4 of 7 ...

Страница 5: ...r this example the max capacity of the 1210AF is 10K so we enter 10000 and then press ENTER Or you can enter the appropriate SEB Output Value as your full scale output value 13 Enter the voltage excitation using buttons and T hen press ENTER 14 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 5 of 7 ...

Страница 6: ...ttons and Then press ENTER For the next display you will be required to enter the mV V for the full scale output of the direction that you are going to use In this example we will pick Tension We see that the value of 4 07906 is the full scale mV V correlating with the max capacity of 10K that we entered earlier 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setu...

Страница 7: ...8a 18b Calibration setup is now complete To return to the main menu press ESC 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 7 of 7 ...

Отзывы: