background image

5

OHM

Range

Resolution

Accuracy

200Ω

0.1Ω

± (1.0% ± 10)

2kΩ

± (1.0% ± 4)

20kΩ

10Ω

200kΩ

100Ω

2MΩ

1kΩ

20MΩ

10kΩ

± (1.0% ± 10)

DWELL ANGLE

Range

Scope (°)

Resolution

Accuracy

4 CYL

0 ~ 90

0.1°

± (2.0% 

± 5)

6 CYL

0 ~ 60

8 CYL

0 ~ 45

TACHO

Range

Scope (°)

Resolution

Accuracy

4 CYL

20,000 

PRM

10 x RPM

± (1.5% 

± 5)

6 CYL
8 CYL

AC & DC VOLTAGE MEASUREMENT

• 

Set the function range switch at the required position.

• 

Connect the black test lead to the "COM" terminal and the red test lead to the "VΩ" 

input terminal.

• 

Connect the test leads to the measuring point and the polarity of the red lead 

connection will be indicated at the same time as the voltage, on the LCD.

• 

If the voltage to be tested is unknown, set the Function Range switch to the 

highest range.

Note: When only figure "1" is displayed, overrange is being indicated and the function 

range switch should to be set to a higher range.

WARNING: Do not attempt to measure voltage above 1000V as there is a danger of 

damaging the internal circuitry.

AC & DC CURRENT MEASUREMENT

• 

Connect the black test lead to the "COM" terminal and the red test lead to the 

"mA" terminal for a maximum of 0.2A.

• 

Set the function range switch at the required position.

• 

Connect the test leads to the measuring points and read the display value. The 

polarity at the red test lead connection will be indicated at the same time as the 

current.