MEASURE TEMPERATURE

• 

Turn the rotary knob to “TEMP”. 

• 

Insert the K-type thermocouple into the input socket of the instrument, with the 

positive end of the thermocouple (red) in the “    ” socket, and the negative end 

(black) in the “COM” socket.

• 

Contact the probe of the thermocouple to the measured object and read the result 

from the main display area on the screen.

• 

Press the “°C/°F” button to switch between celsius and fahrenheit unit.

•  Note

: When the thermocouple is not inserted it will display the temperature of the 

environment.

• 

The instrument is not suitable for measuring the temperature of a rapidly changing 

environment. 

• 

The cold junction compensation of the thermocouple is set in the instrument, 

and it takes a longer time to reach thermal balance with the environmental 

measurements. When measuring without the thermocouple, the instrument 

requires an adjustment period in the environment to obtain accurate readings. 

• 

The instrument uses K-type thermocouple probe.

WARNING

• 

When measuring temperature by thermocouple, the probe of the thermocouple 

cannot contact the charged object, otherwise it may cause the instrument 

damage, electric shock or personal injuries.

• 

Caution: Do not fold the thermocouple wire into an acute angle. Repeatedly 

bending the wire over a period of time may cause the wire to break. 

MEASURE TRANSISTOR

• 

Turn the rotary knob to “hFE”. 

• 

Insert the multi-functional test socket in “10A” and “mA” sockets as 

shown in the figure on the right. 

• 

Insert the transistor in the corresponding socket of the multi-functional 

test socket. 

• 

Read the measurement results on the screen.

• 

Use caution when using the multi-functional, pay attention to the 

direction of the insertion; if it is reversed the test result will be wrong.

WARNING

• 

Do not use the multi-functional test socket in voltage and current measurement, 

otherwise it may cause damage to the instrument, electric shock or personal 

injuries.

AUTO POWER-OFF FUNCTION

• 

If there’s no operation for 15 minutes while the unit is turned on, the instrument will 

enter ‘Sleep’ mode, meaning it will power off automatically to save battery power.

• 

After entering sleep mode, the instrument will turn on if the rotary switch or any 

buttons are pressed. 

• 

If the “FUNC.” button is pressed to turn on the unit, it will cancel the auto power-off 

function.

•  Note

: When USB function is enabled, the instrument will not enter the sleep state.

Содержание D03122

Страница 1: ...1 MODEL D03122 DIGITAL MULTIMETER WITH T RMS USB...

Страница 2: ...9 Measure Resistance 10 Measure Connectivity 10 Measure Capacitance 10 Measure Diode 11 Measure Temperature 11 Measure Transistor 11 Auto Power Off Function 12 Backlight Function 12 Data Hold 12 Gener...

Страница 3: ...ves masks and flame retardant clothes etc to prevent being injured by electric shock due to exposed hazardous live conductors When measuring connect the zero line or the ground line first then connect...

Страница 4: ...6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 1 Auto power off 8 DC indicator 15 Max Min Value Measurement 2 Simulation bar indicating rapid change trend 9 Resistance Frequency Unit 16 USB Interface Open...

Страница 5: ...on selections 1 OFF 2 AC or DC Voltage Press Func to switch 3 AC or DC Voltage mV Press Func to switch 4 Frequency duty ratio Press Hz to switch 5 Resistance diode connectivity capacitance Press Func...

Страница 6: ...perform the following steps In automatic ranging mode press the RANGE button the AUTO symbol will disappear from the display Press the RANGE button to increase the range When the maximum range is reac...

Страница 7: ...n except OFF Press and hold the REL USB button for more than 2 seconds and the instrument will display the USB symbol to show that the data transmission function is enabled Press and hold the REL USB...

Страница 8: ...electric shock or personal injuries Do not apply voltage exceeding DC 1000V or AC 750V RMS between the common terminal and the earth otherwise it may cause damage to the instrument electric shock or...

Страница 9: ...mage to the instrument or equipment please check the fuse before measurement and use the correct input socket MEASURE RESISTANCE Scroll the rotary knob to press the FUNC button and switch to resistanc...

Страница 10: ...the measurement result on the screen WARNING Before measuring resistance connectivity capacitance or diode please turn off the power supply and discharge all the high voltage capacitors otherwise it...

Страница 11: ...rwise it may cause the instrument damage electric shock or personal injuries Caution Do not fold the thermocouple wire into an acute angle Repeatedly bending the wire over a period of time may cause t...

Страница 12: ...the use of the backlight then replace the battery immediately DATA HOLD In the process of measurement if a held reading is required press the button the displayed value on the screen will be held Pres...

Страница 13: ...oint the reading will stabilize 11 Power Requirement 4 x 1 5V AA batteries Dimensions 204x94x57mm LxWxH Weight About 410g inluding batteries AC VOLTAGE Range Resolution Accuracy Input impedance 10M 60...

Страница 14: ...easurement the instrument should be cooled down after twice of the measurement time then enter small current measurement RESISTANCE Range Resolution Accuracy Open circuit voltage about 0 25V Overload...

Страница 15: ...range range the input current should be increased with the increase of the measured frequency Input protection A mA range F600mA 1000V ultra speed fuse 10A range F10A 1000V ultra speed fuse Note Comp...

Страница 16: ...r Remove the old batteries and replace with new batteries of the same specification Ensure that the batteries are inserted correctly observing the plus and minus marks on both the batteries and compar...

Отзывы: