7 | Checking / setting parameters with D‑ESI 100
128
D-R 290
Setting path:
Site map > »Device (e.g. D‑R 290 ) > >
Common parameters > Variables
setting > [
type code
] >
Set the output variable (type code) of the measurement channel
Left click on "device" in the site map.
Left click on for the desired device in the device list.
Under
Common parameters
in the list field, after "
Select group"
, select "
Var
iable setting"
.
Left click on the "Code table" symbol
in the "Type code in the measurement
channel" line (#1 … #4) for which you wish to change the setting.
Left click on the radio button (column On/Off) in the line which contains the desired
output variable (type).
●
Left click on the "
Close
" button, to close the window
without loading the new value
to the measuring head.
●
Left click on the "
Send
" button to load the new value to the temporary memory of
the measuring head and then close the window.
The new setting will immediately be used until the device is next restarted. After a re
start, the old setting will be used. In order to use the new setting permanently (even
after a restart), the settings must be saved before the next restart.
To do this, left click on the "Save parameters" button. (The new setting is not used
permanently until it has been saved.)
Available type codes for the D-R 290
The following type codes can be set for the D-R 290:
Type
Unit
Description
Dust
mg/m³
Dust content, calculated from the optical density and the
calibration parameters
4.2.7 Calculation of the dust concentration [
and
Dust (standard)
mg/Nm³
Dust content converted to standard cubic metres
7.7.10 Temperature compensation [
Transmission
% T
Optical transmission in the measuring path. The display
shows the transmission which the light encounters in tra
versing the measuring path once.
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Save parameters
7.5.2.1
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...