- 13 -
Press
button
LED light Blink
Displaying Ramp up
rate to T2 and blink
Use up/down and left/right
arrow to change ramp rate.
Input 0 will skip this step
Press
button
LED light Blink
Displaying hold time
at temp. T2
Use up/down arrow to
change time for staying at
T2. Input 0 will skip this step
Press
button
LED light Blink
Displaying T3
temperature and blink
Use up/down and left/right
arrow to change T3
temperature. Input 0 will skip T3
Press
button
LED light Blink
Displaying Ramp
rate to T3 and blink
Use up/down and left/right
arrow to change ramp rate.
Input 0 will skip this step
Press
button
LED light Blink
Displaying hold time
at temp. T3
Use up/down arrow to
change time for staying at
T3. Input 0 will skip this step
Set up ramp down
rate
Use up/down arrow to change time
to input hold time at T3
If it is “0”, heater off after finishing T3 progress.
Schedule will include cool down time till 250
o
C with 25
o
C/min. rate.
If it is “1”, schedule will progress with controlling temperature down.
Heater will turn off after cool down till 250
o
C with 15
o
C/min. rate.
If it is “2”, staying at maximum temp. for 10 sec. with alarm.