11. Repeat steps 6-10 for channel 2.
12.
For channel 3, change the ‘unconverted’ signal to
Voltage
in the signal properties
window.*
13.
Enter the battery voltage levels and calibrations found on the sterile package and select
OK
.
*Please note that for A.R.T. versions 2.2 and earlier, the ‘voltage’ signal type is not
available. Channel 3 should be left as ‘unconverted’ and the battery voltage data will not
be usable.
New calibration
temperatures from
the previous step
appear here
Note: The calibration values
shown are examples. Use
those found on each device’s
sterile package.