Series PC Manual V 1.0 - Rev. 3
– 03/01/01
DS Europe
29
Open the transducer lid when
transducer is off.
Power the transducer:
Green
LED
will start blinking.
Yellow LED
will show
transducer status
Cursor 0 or
Cursor 1 ?
Cursor 0
Press push button
A
for approx. 2 sec.
till
Yellow LED
is off.
Press push button
B
for approx. 2 sec.
till
Yellow LED
is off.
Analog output of A/D1 is fluctuating
between
0V
and
2,5V
.
Move cursor onto
ZERO
position.
Press push button
A
and
Yellow LED
will blink once for
1 sec
. confirming
the
ZERO
point memorization.
Analog output of A/D1 is fluctuating
between
2,5V
and
5V
.
Move cursor to
FULL SCALE
point.
Press push button
A
and
Yellow LED
will blink for
1 sec
. confirming the
FULL SCALE
point memorization
IS YELLOW
LED ALSWAYS
ON ?
Analog output of A/D1 is fluctuating
between
5 V
and
7.5 V.
Move cursor onto
ZERO
position.
Press push button
B
and
Yellow LED
will blink once for
1 sec
. confirming
the
ZERO
point memorization
Analog output of A/D1 is fluctuating
between
7
,
5 V
and
10V.
Move cursor to
FULL SCALE
point.
Press push button
B
and
Yellow LED
will blink for
1 sec
. confirming the
FULL SCALE
point memorization
Power off and then on the
transducer in order to reset It
and to activate the just done
calibration.
IS YELLOW
LED ALSWAYS
ON ?
NO
NO
Note: if Yellow LED remains on
there has been a calibration error
and probably zero and full scale
points are the same point. Repeat
the calibration procedure
mantaining at least a 10% Full
Scale distance between zero and
Full Scale points.
Cursor 1
YES