27
12) TESTS FOR THE APPLICATION OF THE CE MARK
12.2 AMPLIFIED TRANSDUCER:
They are subjected to CE mark.
Those one without internal amplifier are considered “passive components” and therefore
any certification is necessary.
12.2 TEST CENTER AND LIST OF THE TESTS:
The tests have been done at a Test Centre “competent body” internationally approved.
For the compatibility tests have been chosen the standards EN 50081-2 (for emission)
and EN 50082-2 (for immunity), the most severe for industrial applications.
12.2.1 Normative reference: EN 55011: Test category: Emission; Port: enclosure; Type of
test: radiated interference field strength; Frequency range: 30 to 1000 MHz.
12.2.2 Normative reference: ENV 50140: Port: enclosure; Test category: Immunity; Type
of test: radiated radio-frequency, electro-magnetic field; Frequency range: 80 to
1000 MHz; Test level: 10 V/m.
12.2.3 Normative reference: ENV 50141: Port: DC I/O power port; Test category:
Immunity; Type of test: RF common mode; Frequency range: 0,15 to 80
MHz; Test level: 10 V.
12.2.4 Normative reference: ENV 50141: Port: signal lines; Test category: Immunity; Type
of test: RF common mode; Frequency range: 0,15 to 80 MHz;
Test level: 10 V.
12.2.5 Normative reference: EN 61000-4-4: Port: DC I/O power port; Test category:
Immunity; Type of test: Fast transient (burst). Common mode;
Test level: 3.
12.2.6 Normative reference: EN 61000-4-4: Port: signal lines; Test category: Immunity;
Type of test: Fast transient (burst) common mode; Test level: 3.
12.2.7 Normative reference: EN 61000-4-2: Port: Enclosure; Test category: Immunity;
Type of test: ESD; Test level:
±
4kV.
12.2.8 Normative reference: EN 61000-4-8: Port: Enclosure; Test category: Immunity;
Type of test: Power frequency magnetic field. Frequency range: 50 Hz;
Test level: 4.
12.2.8 Normative reference: ENV 50204: Port: Enclosure; Test category: Immunity;
Type of test: Radio frequency electromagnetic field. Pulse modulated: 900 + 5
MHz; Test level: 3.
Notes:
1)
The tests have given positive results.
2)
The test Laboratory and the EC normatives compel reserve on the paper-works and on the
data of the tests; they neither can be advertised nor photocopied and they remain at
disposal only to the Competent Authorities in our Factory.
Содержание 500QD Series
Страница 23: ...23 ...