DRUCKER DIAGNOSTICS 
HORIZON 24 SERVICE MANUAL 

SM016 

REV A 

 

5.

 

SPECIFICATIONS 

Maximum Speed  

3400 RPM 

Maximum RCF 

2000 

Dimensions (in) 

9.0 (H) x 15 (W) x 17 (L) 

Environmental Operating Range 

16-32 deg C 

Typical Noise Level (At Maximum Speed) 

< 59 dB A 

 

 

Electrical Rating 

95-253 VAC 

 

6.

 

TROUBLESHOOTING 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

The lid does not open. 

No Power 

Check removable line cord or wall outlet. 

Lid knob is ajar 

Rotate the lid knob fully clockwise before pressing the 

‘UNLOCK’ button

 

Lid lock is active (Unlock timed out) 

Press the ‘UNLOCK’ button to de

-activate the lid 

Lid tray is unplugged from PCB or 
defective 

Requires service 

PCB is damaged 

Requires service 

 

To gain access to the rotor - Remove the label directly in 
front of the lid knob and slide the lid latch lever toward 
the front of the centrifuge. This will unlock the lid.  

 

 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

Excessive vibration  

Rotor improperly loaded 

Load equally filled tubes symmetrically in the rotor. All 
carriers and/or tube holders must be present in the 
rotor, whether loaded, or empty. 

Debris lodged within the rotor or 
tube carriers 

Carefully inspect all rotor pockets, tube holders and 
crevasses for debris.  

Centrifuge housing is loose 

Requires service 

Missing/damaged feet 

Requires service 

Motor failure 

Requires service 

Rotor damaged 

Replacement required 

 

 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

Rotor does not spin 

No Power 

Check removable line cord 

No Power 

Check circuit breaker on underside of centrifuge. 

No Power 

Check wall outlet 

Lid not properly latched 

Press down firmly on lid and rotate lid knob fully 
clockwise. 

Internal connection failure 

Requires service 

PCB failure 

Requires service  

Motor Failure 

Requires service  

 

 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

Clicking noise during braking 

Rotor is loose 

Tighten rotor screw per section 7-2 

 

 

 

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

Whistling noise while running 

Debris in air intake / exhaust ports  

Remove power before clearing debris. 

Gasket failure 

Requires service 

 

 

Содержание Horizon 24

Страница 1: ...DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL SM016 REV A 1 200 Shadylane Drive Philipsburg PA 16866 www druckerdiagnostics com Service Manual Horizon 24 Centrifuge ...

Страница 2: ...ICE MANUAL SM016 REV A 2 Contents 1 PREFACE 3 2 INTENDED USE 3 3 GENERAL DESCRIPTION OF MAJOR COMPONENTS 3 4 WARRANTY INFORMATION 3 5 SPECIFICATIONS 4 6 TROUBLESHOOTING 4 7 SERVICE INSTRUCTIONS 5 8 WIRING DIAGRAM 11 9 SPARE PARTS LIST 12 ...

Страница 3: ...ntainers filled with chemicals non flammable non explosive non volatile and non highly reactive only environmental samples and other non human body samples This centrifuge should be operated by trained personnel only This centrifuge is an IVD accessory and therefore subject to the former EU IVD Directive 98 79 EC and the current EU IVD Regulation 2017 746 Any use other than those intended by the M...

Страница 4: ...Rotor improperly loaded Load equally filled tubes symmetrically in the rotor All carriers and or tube holders must be present in the rotor whether loaded or empty Debris lodged within the rotor or tube carriers Carefully inspect all rotor pockets tube holders and crevasses for debris Centrifuge housing is loose Requires service Missing damaged feet Requires service Motor failure Requires service R...

Страница 5: ...ilure Replace PCB 7 SERVICE INSTRUCTIONS 7 1 Cleaning The cabinet rotor and accessories can be cleaned using soap and water isopropyl alcohol or a 10 5500 PPM bleach solution Apply cleaning solutions with a towel or cloth Do not submerge the centrifuge in water or other cleaning solutions as this will cause damage and void the warranty Under no circumstances should any of the following be used TBQ...

Страница 6: ...n place with two 6 Nylok nuts Use a 5 16 nut driver to remove the nuts The tray will slide off the threaded studs To install the lid tray slide it onto the studs and secure it with two 6 Nylok nuts hand tight Complete the installation by gently plugging the wire harness into the PCB header J2 The header and connector are keyed for proper orientation 7 8 Replacing the PCB The PCB is accessible once...

Страница 7: ...ce with six 8 screws Remove them with a 2 Phillips driver Lift the base assembly off the guard bowl and set it aside The motor assembly is held in place with 3 8 Nylok nuts and fender washers Remove the nuts with an 11 32 nut driver Remove the three fender washers Lift the motor assembly up and off the studs The motor is secured to the mounting plate with 4 8 screws Use a 2 Philips driver to remov...

Страница 8: ...DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL SM016 REV A 8 Fig 3 Fig 4 Fig 5 Insert the Elastomer bushings into the motor mount plate as shown fig 4 Push the grommets onto the bushings as shown fig 5 ...

Страница 9: ... of the guard bowl splice fig 7 Install one fender washer onto each stud fig 8 Finish the assembly with one Nylok nut on each stud and tighten to 13 in lbs Fig 9 Fig 10 Place the base assembly onto the guard bowl assembly 02 003 0 0052 with the guard bowl splice positioned as shown fig 9 Feed the motor wire harness into the base passthrough so that the wire harness is not pinched between the base ...

Страница 10: ...ed to 10 in lbs fig 12 7 11 Power Connections and Final Assembly Plug the power harness into J17 Plug the motor harness into J10 Hall Sensors and J14 Motor power Plug the lid lock harness into J2 Plug the cycle counter harness into J3 Carefully place the cabinet onto the base taking care not to pinch any wires between the two Complete the assembly by replacing the eight 8 screws with a 3 32 hex ke...

Страница 11: ...DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL SM016 REV A 11 8 WIRING DIAGRAM ...

Страница 12: ...DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL SM016 REV A 12 9 SPARE PARTS LIST ...

Отзывы: