Setting A
Setting B
Setting C
RPM
1,500
2,500
1,500
Time
5 min
4 min
4 min
R
EVIEW
C
YCLE
T
IME AND
S
PEED
S
ETTINGS
Factory programmed cycles are shown on the rear of the centrifuge, on the Factory Set Cycles label. To review
current settings, follow this procedure:
o
The lid must be open to review the selected cycle time and speed.
o
Press and hold the START button until you hear a beep.
o
Release the START button. The centrifuge will beep and the LED light will flash once for each minute of run time
in the current cycle. 10 beeps / flashes equal 10 minutes of run time. Run time starts when the rotor reaches 90%
of desired speed and stops when the rotor starts decelerating.
o
Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100 rpm in
the current cycle. 38 beeps / flashes equal 38 x 100 or 3,800 Revolutions per Minutes (RPM).
o
The centrifuge will automatically revert to normal mode at the end.
C
HANGING
C
YCLE
T
IME AND
S
PEED
S
ETTINGS
o
The lid must be open to change the selected cycle time and speed.
o
Select the cycle you wish to change.
o
Press and hold the START and UNLOCK buttons together until the LED light flashes.
o
Press the START button for each minute of run time.
o
Move to speed setting mode by
pressing the UNLOCK button.
o
Press the START button once for each 100 rpm.
o
Press the UNLOCK button to exit setting mode.