Drucker Diagnostics Horizon 12 Скачать руководство пользователя страница 6

DRUCKER DIAGNOSTICS 
HORIZON 12 SERVICE MANUAL 

SM015 

REV A 

 

7.6.

 

Removing the Cabinet (Upper Housing) 

 

There are eight screws that fasten the centrifuge cabinet to the base. 

 

Begin by unplugging the centrifuge. 

 

Use a 3/32 hex key to remove the cabinet screws (three on left and right sides, two 
in the rear) 

 

The cabinet control panel is attached to the base internally with cable harnesses. 
Be careful not to stress the cables when removing the cabinet. 

 

Stand directly in front of the centrifuge and lift the cabinet straight up and off the 
base, setting it down on its right side. 

 

Gently remove the power supply and motor harnesses from the PCB.  

7.7.

 

Replacing the Lid Lock Assembly 

 

The lid tray assembly is accessible once the cabinet has been removed. 

 

Gently remove the lid tray wire harness from the PCB. 

 

The lid tray assembly is held in place with two #6 Nylok nuts. Use a 5/16” nut driver 

to remove the nuts. The tray will slide off the threaded studs. 

 

To install the lid tray, slide it onto the studs, and secure it with two #6 Nylok nuts, 
hand tight. 

 

Complete the installation by gently plugging the wire harness into the PCB header 

‘J2’. The header and connector are keyed for proper orientation.

 

7.8.

 

Replacing the PCB 

 

The PCB is accessible once the cabinet has been removed. Make certain that all 
wire harnesses have been disconnected. Use standard precautions for handling 
static sensitive components. 

 

The PCB is held in place with six #6 Nylok nuts and plastic insulator washers. Use a 

5/16” nut driver to remove the nuts. The PCB and washers are now free to slide

 off 

the threaded studs. 

 

Beneath the PCB are six plastic standoffs. If they are crushed, replace them before 
reassembly. 

 

To install the PCB, make certain that six plastic standoffs are present on the control 
panel studs. 

 

Slide the board onto the studs 

 

Install six plastic washers onto the studs securing the circuit card. 

 

The six nuts should be tightened only until the buttons on the control panel click 
when pressed 

 do not over tighten. Each nut sets the height of a separate control 

panel button. Too low, and they cannot be actuated, too high and they jam against 

the front panel label and are always in the ‘ON’ state. Tighten each nut a ¼ turn in 

succession, trying each corresponding button as you go. 

 

Once each control panel button clicks when pressed, installation is complete. 

 

 

Содержание Horizon 12

Страница 1: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 1 200 Shadylane Drive Philipsburg PA 16866 www druckerdiagnostics com Service Manual Horizon 12 Centrifuge...

Страница 2: ...CE MANUAL SM015 REV A 2 Contents 1 PREFACE 3 2 INTENDED USE 3 3 GENERAL DESCRIPTION OF MAJOR COMPONENTS 3 4 WARRANTY INFORMATION 3 5 SPECIFICATIONS 4 6 TROUBLESHOOTING 4 7 SERVICE INSTRUCTIONS 5 8 WIR...

Страница 3: ...tainers filled with chemicals non flammable non explosive non volatile and non highly reactive only environmental samples and other non human body samples This centrifuge should be operated by trained...

Страница 4: ...carriers Carefully inspect all rotor pockets tube holders and crevasses for debris Centrifuge housing is loose Requires service Missing damaged feet Requires service Motor failure Requires service Ro...

Страница 5: ...he warranty Under no circumstances should any of the following be used TBQ Germicidal Products Fully Partially Halogenated Hydrocarbons Ketones or Esters The use of any chemicals not prescribed by the...

Страница 6: ...tion by gently plugging the wire harness into the PCB header J2 The header and connector are keyed for proper orientation 7 8 Replacing the PCB The PCB is accessible once the cabinet has been removed...

Страница 7: ...t it aside The motor assembly is held in place with 3 8 Nylok nuts and fender washers Remove the nuts with an 11 32 nut driver Remove the three fender washers Lift the motor assembly up and off the st...

Страница 8: ...s as shown fig 5 Fig 6 Fig 7 Fig 8 Place one Aluminum spacer 30 000112 onto each stud as shown fig 6 Install the motor mount assembly onto the guard bowl studs so that the motor wire tiedown is positi...

Страница 9: ...plice positioned as shown fig 9 Feed the motor wire harness into the base passthrough so that the wire harness is not pinched between the base and the guard bowl fig 10 Fig 11 Fig 12 Secure the base w...

Страница 10: ...rness into J17 Plug the motor harness into J10 Hall Sensors and J14 Motor power Plug the lid lock harness into J2 Plug the cycle counter harness into J3 Carefully place the cabinet onto the base takin...

Страница 11: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 11 8 WIRING DIAGRAM...

Страница 12: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 12 9 SPARE PARTS LIST...

Отзывы: