DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL
SM015
REV A
1
200 Shadylane Drive
Philipsburg, PA 16866
www.druckerdiagnostics.com
Service Manual
Horizon 12 Centrifuge
Страница 1: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 1 200 Shadylane Drive Philipsburg PA 16866 www druckerdiagnostics com Service Manual Horizon 12 Centrifuge...
Страница 2: ...CE MANUAL SM015 REV A 2 Contents 1 PREFACE 3 2 INTENDED USE 3 3 GENERAL DESCRIPTION OF MAJOR COMPONENTS 3 4 WARRANTY INFORMATION 3 5 SPECIFICATIONS 4 6 TROUBLESHOOTING 4 7 SERVICE INSTRUCTIONS 5 8 WIR...
Страница 3: ...tainers filled with chemicals non flammable non explosive non volatile and non highly reactive only environmental samples and other non human body samples This centrifuge should be operated by trained...
Страница 4: ...carriers Carefully inspect all rotor pockets tube holders and crevasses for debris Centrifuge housing is loose Requires service Missing damaged feet Requires service Motor failure Requires service Ro...
Страница 5: ...he warranty Under no circumstances should any of the following be used TBQ Germicidal Products Fully Partially Halogenated Hydrocarbons Ketones or Esters The use of any chemicals not prescribed by the...
Страница 6: ...tion by gently plugging the wire harness into the PCB header J2 The header and connector are keyed for proper orientation 7 8 Replacing the PCB The PCB is accessible once the cabinet has been removed...
Страница 7: ...t it aside The motor assembly is held in place with 3 8 Nylok nuts and fender washers Remove the nuts with an 11 32 nut driver Remove the three fender washers Lift the motor assembly up and off the st...
Страница 8: ...s as shown fig 5 Fig 6 Fig 7 Fig 8 Place one Aluminum spacer 30 000112 onto each stud as shown fig 6 Install the motor mount assembly onto the guard bowl studs so that the motor wire tiedown is positi...
Страница 9: ...plice positioned as shown fig 9 Feed the motor wire harness into the base passthrough so that the wire harness is not pinched between the base and the guard bowl fig 10 Fig 11 Fig 12 Secure the base w...
Страница 10: ...rness into J17 Plug the motor harness into J10 Hall Sensors and J14 Motor power Plug the lid lock harness into J2 Plug the cycle counter harness into J3 Carefully place the cabinet onto the base takin...
Страница 11: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 11 8 WIRING DIAGRAM...
Страница 12: ...DRUCKER DIAGNOSTICS HORIZON 12 SERVICE MANUAL SM015 REV A 12 9 SPARE PARTS LIST...