The centrifuge control panel is attached to the base internally with cable 
harnesses. Be careful not to stress the cables when removing the cabinet. 

 

Lift the cabinet straight up and off the base, setting it down on its front side. 

 

Gently remove the combination power supply / motor harness from the control 
board.  

6.7.

 

Replacing the Lid Lock Assembly 

 

The lid locking tray assembly is accessible once the cabinet has been removed. 

 

Gently remove the lid tray wire harness from the PCB. 

 

The lid tray assembly is held in place with two #6 Nylok nuts. Use a 

5/16”

 nut driver 

to remove the nuts. The tray will slide off the threaded studs. 

 

To install the lid tray, slide it onto the studs, and secure it with two #6 Nylok nuts, 
hand tight. 

 

Complete the installation by gently plugging the wire harness into the PCB header 

‘J1’.

 

6.8.

 

Replacing the Control Board 

 

The control board is part of the front panel assembly and is accessible once the 
cabinet has been removed. Make certain that all wire harnesses have been 
disconnected.  

 

The control board has capacitors that will remain charged for a period after the 
centrifuge is unplugged. Make certain to use standard precautions for handling 
potentially charged capacitors when working with the control board. 

 

The front panel assembly is held in place with ten locking tabs (fig 1). 

 

Push the tabs inward to release the assembly. 

 

To install the front panel assembly, push it into the opening in the cabinet until it 
snaps into place.  

 

Figure 1 

 

 

Содержание 642M

Страница 1: ...DRUCKER DIAGNOSTICS 642M SERVICE MANUAL SM023 REV C 1 200 Shadylane Drive Philipsburg PA 16866 www druckerdiagnostics com Service Manual 642M CENTRIFUGE...

Страница 2: ...SERVICE MANUAL SM023 REV C 2 Contents 1 PREFACE 3 2 GENERAL DESCRIPTION OF MAJOR COMPONENTS 3 3 WARRANTY INFORMATION 3 4 SPECIFICATIONS 3 5 TROUBLESHOOTING 4 6 SERVICE INSTRUCTIONS 5 7 WIRING DIAGRAM...

Страница 3: ...the microcontroller based control center of the centrifuge All control signals are generated in the control board 2 3 Lid Locking Tray Assembly The lid tray assembly contains a solenoid and limit swi...

Страница 4: ...e carriers Carefully inspect all rotor pockets tube holders and crevasses for debris and clean thoroughly Centrifuge housing is loose Tighten or replace screws Missing damaged feet Replace feet Motor...

Страница 5: ...thumb is one full thread showing above the nut is a good indicator of proper installation 6 3 Maintaining the Rotor Keep the rotor clean any corrosive materials must not be allowed contact with the r...

Страница 6: ...ff the threaded studs To install the lid tray slide it onto the studs and secure it with two 6 Nylok nuts hand tight Complete the installation by gently plugging the wire harness into the PCB header J...

Страница 7: ...acing the motor Flip the guard bowl upside down and locate the notch on the bottom Install the motor into the guard bowl with the wires facing in the same direction as the notch Holding the motor in p...

Страница 8: ...nd Final Assembly Connect all wiring harnesses to the control board as shown in the wiring diagram Carefully place the cabinet onto the base taking care not to pinch any wires between the two Complete...

Страница 9: ...DRUCKER DIAGNOSTICS 642M SERVICE MANUAL SM023 REV C 9 7 WIRING DIAGRAM...

Страница 10: ...DRUCKER DIAGNOSTICS 642M SERVICE MANUAL SM023 REV C 10 8 SPARE PARTS LIST...

Отзывы: