2 | PV 211 Instruction Manual–English

2.1  Hose/Adaptors (Figure

 

A1)

To attach a hose

 

(5) and adapter

 

(6) to a connector

 

(4), 

turn the knurled nut on the connector counterclockwise.

2.2  Pressure Release Valve (Figure

 

A1)

Use this

 

(1) to reduce or release the pressure in the 

system. The amount of turn sets the rate to release the 

pressure. Only minimum force is necessary to seal the 

system.

2.3  Fine Adjust Vernier (Figure

 

A1)

To make accurate adjustments to the pressure, turn the 

fine adjust vernier

 

(2) clockwise to increase the pressure or 

counterclockwise to decrease the pressure.

2.4  Pressure/Vacuum Selector (Figure

 

A1)

Before you change the mode, make sure that the pressure 

release valve

 

(1) is open. To change the mode, use an 

applicable tool to push the selector

 

(3) to the side specified 

on the label.

2.5  Overpressure Protection (Figure

 

A1)

To adjust the maximum output pressure of the system, turn 

the nuts

 

(8) to increase or decrease the stroke length.

3. Operation

3.1  Calibration/Comparison against an 

Analog Gauge

1. Connect a reference instrument to the pressure port at 

the top of the pump

 

(9).

2. Connect the instrument under test to the connection at 

the end of the flexible hose

 

(5).

Note: 

The maximum torque for the adaptors is 15

 

Nm 

(11

 

ft lb).

3. Set the fine adjust vernier

 

(2) to the midpoint of its 

travel: 
Turn it fully clockwise, then four to six turns 

counterclockwise.

4. Close the pressure release valve

 

(1): 

Turn it fully clockwise, and tighten to seal.

5. Operate the scissor handles

 

(7) until the pressure is 

almost correct. To get the maximum pressure output, 

squeeze the handles fully together on each stroke.

6. To adjust the pressure to the correct value, turn the fine 

adjust vernier

 

(2) clockwise to increase the pressure or 

counterclockwise to decrease the pressure.

Note: 

Initially, small pressure changes can occur 

(thermodynamic effects, the seals settle, the hoses 

expand). The pressure will stabilize after a short time.

7. You can also decrease the pressure by careful 

operation of the pressure release valve

 

(1).

8. To get a vacuum, set the pressure/vacuum selector

 

(3) 

to the vacuum position (as indicated on the pump label) 

then use the same procedure as above.

Note: 

Before you change the mode, release the 

pressure.

3.2  Operation with a High-Resolution 

Pressure Calibrator

Because the resolution is better, it is possible you will see 

more of the small pressure changes identified in step 6 

above.

Note: 

On very high resolutions such as 1

 

mbar or 

0.1

 

inH

2

O, small movement of the hose may result in 

noticeable pressure changes.

4. Fault Finding

1. If the system appears to lose pressure, repeat the 

above procedure. Make sure: there is no damage to the 

seals, the adapters are tightened sufficiently, the 

pressure release valve

 

(1) is tightened sufficiently to 

seal.

2. Do not try to tighten the connections between the pump 

and the fine adjust vernier

 

(2) or between the pump and 

the quick-fit connectors

 

(4). These are factory set and 

changes can cause damage to the sealed joints.

3. During leak tests, small air movements (in or out) are 

possible around the pressure/vacuum selector

 

(3). 

This is normal.

4. If the PV

 

211 has not been used for a period of time, it 

can be difficult to operate on the first stroke. It will 

become free after this.

5. For seal replacement refer to the service kit 

instructions.

If, for any reason, a fault occurs within the pump, it is 

recommended that the equipment be returned to an 

appointed agent.

INFORMATION  

To prevent damage, DO 

NOT use force to turn the fine adjust 

vernier

 

(2) farther than the red line indicator 

on the thread. See item 11 in Figure

 

A2.

WARNING  

Uncontrolled release of high 

pressure is dangerous and may cause 

damage to equipment. Because the internal 

pressure can get very high during operation, 

make sure that all the connections are made 

correctly.
DO NOT ignore the maximum operating 

pressure specified on the pump label.
Before you connect a pressure component to 

the PV

 

211, make sure that it is isolated from 

the pressure supply and release the internal 

pressure slowly. DO NOT connect the pump to 

an external pressure source.

INFORMATION  

To prevent damage, DO 

NOT use force to turn the fine adjust 

vernier

 

(2) farther than the red line indicator 

on the thread. See item 11 in Figure

 

A2.