![Druck InstruMetrics ADTS542F Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/druck/instrumetrics-adts542f/instrumetrics-adts542f_safety-and-installation-manual_2533090006.webp)
Copyright 2014 Baker Hughes Company.
4 | ADTS Safety and Installation Guide–English
Figure
3: ADTS Touch Umbilical Cable Connector
The ADTS Touch can contain one Lithium-ion battery which is housed in the battery compartment
of the ADTS Touch.
2.1 Replacing the Battery
Figure
4: ADTS Touch Battery Compartment
Note:
Only use batteries that have been supplied by or purchased from Druck.
Prior to replacing the battery, switch the ADTS Touch off. If the ADTS Touch is connected to the
umbilical cable or Docking connector, disconnect prior to removing the battery cover plate.
1. Unscrew the two thumbscrews (5) of the battery compartment cover plate and remove the
cover plate.
2. Use the tab attached to the battery to slide the battery out of the battery compartment.
3. Fit a new or re-charged battery into the battery compartment, observing the keyed
orientation.
CAUTION
When the battery compartment cover plate is removed and the
ADTS touch is tilted with the battery compartment facing downwards, the battery
can fall out of the ADTS touch causing damage to the battery.
4
5
5
Содержание InstruMetrics ADTS542F
Страница 1: ...Druck com ADTS 542F 552F 553F 554F Air Data Test Set Safety and Installation Guide ...
Страница 2: ......