Druck InstruMetrics ADTS542F Скачать руководство пользователя страница 1

Druck.com

ADTS 

542F/552F/553F/554F

Air Data Test Set

Safety and Installation Guide

Содержание InstruMetrics ADTS542F

Страница 1: ...Druck com ADTS 542F 552F 553F 554F Air Data Test Set Safety and Installation Guide ...

Страница 2: ......

Страница 3: ...ents for the ADTS refer to User Manual K0553 For details of available manufacturer specified parts refer to the current revision of the product data sheet Service maintenance and repair will be carried out at Druck authorized service centers 1 1 Markings and Symbols For the list of Druck approved service centers go to Druck com Symbol Description This equipment meets the requirements of all releva...

Страница 4: ...g Capacity fuses 8 see Figure 6 located in the holders and mounted on the front panel protect the unit The fuses are connected in the live and neutral supply circuit and are rated at T5AH 250V 1 7 Replacing a Fuse 1 Disconnect the electrical power supply 2 Unscrew the fuse holder cap from the front panel and remove the fuse 3 Fit a new fuse of the correct size and rating into the fuse holder cap 4...

Страница 5: ... powered on ADTS Using an umbilical cable connected to a powered on ADTS Powered from its own battery pack Powered from a universal power supply unit Figure 1 ADTS Touch Docking Connector The ADTS Touch is equipped with an On Off switch 2 and USB connectors for connection to a PC or mass storage device 3 Figure 2 Touch On Off Switch and USB Connectors Connector 4 allows connection between the base...

Страница 6: ...ch off If the ADTS Touch is connected to the umbilical cable or Docking connector disconnect prior to removing the battery cover plate 1 Unscrew the two thumbscrews 5 of the battery compartment cover plate and remove the cover plate 2 Use the tab attached to the battery to slide the battery out of the battery compartment 3 Fit a new or re charged battery into the battery compartment observing the ...

Страница 7: ... fire Subject the battery to excessive physical shock or vibration outside the specified limits for the ADTS Use a battery that appears to have suffered abuse Charge with an unspecified charger or specified charger that has been modified This can cause breakdown of the battery or swelling and rupturing Use a battery in an appliance or purpose for which it was not intended The battery unit is seale...

Страница 8: ... Installation Guide English Using a universal power supply unit connected to the ADTS Touch using an adapter Figure 5 Universal Power Supply Unit and Adapter The ADTS Touch universal power supply and the desktop charger are both rated for indoor use only ...

Страница 9: ... Power Up Test see Figure 6 1 Fit Pitot Pt 1 and Static Ps 2 port blanks 2 Ensure that the test set is connected to an external power supply and the supply is switched on 3 Use the On Standby Switch 3 on the front of the test set to switch the test set on The test set performs a self test resulting in a test set status indication 4a Pass or Fault 1 3 6 2 4 5 7 8 ...

Страница 10: ...eless connection present Blue Wired connection and Bluetooth option enabled Flashing blue Fast flash enabled and visible for pairing available for 5 minutes after power up Slow blink enabled but not visible for pairing Bluetooth option disabled Wireless LED off Initialization fault Red Only applicable to base unit software variant DK0467 Note The fast flash state also occurs when switching the ON ...

Страница 11: ...cessories on the datasheet are used for connection to the Pitot Pt 1 and Static Ps 2 output ports of the ADTS The Pitot Pt and Static Ps outputs can then be connected to the corresponding Pitot and Static connectors on the aircraft or aeronautical instruments The maximum Pt and Ps output pressure is 2 bar absolute The following table details the possible connector specifications Mating female part...

Страница 12: ...nt ADTS542F 5 C to 50 C 32 to 122 F ADTS552F ADTS553F 5 C to 50 C 32 to 122 F ADTS554F Operating humidity 5 to 95 RH condensing Indoor Outdoor non continuously use ADTS Touch universal power supply Operating temperature 10 C to 50 C 50 to 122 F Operating humidity 5 to 95 RH non condensing Power supply Universal input Mains Power Supply Class 1 equipment requiring a protective earth connection TXAH...

Страница 13: ...y at least two people ADTS5xxF Weight ADTS542F 14kg 31lb without battery estimated as 20kg 44lb with battery ADTS552F 22kg 48 5lb ADTS553F 23 5kg 52lb ADTS554F 24 5kg 54lb Height 300mm Depth ADTS542F 320mm front to rear ADTS55xF 425mm front to rear Width ADTS542F 420mm left to right ADTS55xF 525mm left to right ADTS Touch Weight 1kg 2 2lb 1 4kg 3lb with battery Height 58mm Depth 154mm Width 224mm ...

Страница 14: ... product label for the ADTS Touch can be found on the rear of the ADTS Touch Labeling information consists of 1 Company logo and product name 2 Part number 3 Serial number 4 Date of manufacture DOM 5 Name and address of manufacturer Figure 9 Product Labels ADTS542F PART No ADTS542F SERIAL No DOM Mmm YYYY DRUCK LIMITED LE6 0FH UK ADTS TOUCH PART No ADTSTOUCH SERIAL No DOM Mmm YYYY DRUCK LIMITED LE6...

Страница 15: ... bakerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 31 334678950 nl sensing sales bakerhughes com Moscow 7 915 3161487 aleksey khamov bakerhughes com UAE USA Abu Dhabi 971 528007351 suhel aboobacker bakerhughes com Boston 1 800 833 9438 custcarebosto...

Отзывы: