Insert a new, sterile lancet and 
carefully twist off its protective cap.

Cock Droplet Genteel by pushing in the 
Push Cap, taking care not to hold the 
Activation Button while priming (you will 
hear a click if Droplet Genteel is cocked 
correctly).

Select area to be lanced and wipe with an alcohol swab 
and/or tissue.

Remove the Nozzle by twisting 
counterclockwise, then pull it off.

DO NOT HOLD ONTO THE ACTIVATION BUTTON 

WHILE COCKING DROPLET GENTEEL, AS THIS 

WILL DISENGAGE THE LOCKING MECHANISM 

AND DROPLET GENTEEL WILL NOT COCK.

Testing from the fingertip

Cock Droplet Genteel by pushing in the 
Push Cap at the back of the device.

Twist and take off the Nozzle.

Take out the lancet and place it 
into the sharps container.

13

14
15

7

Installing the Contact Tip onto the Nozzle

Place the Contact Tip on a clean, hard, flat surface, 
like a tabletop.

There are two sides to the Contact Tip—the smooth side that 
goes against your skin and the side with the grooves that goes 
onto the Nozzle. Make sure the side with the groove is facing up. 
Use a slight back and forth (rotating) motion to gently push the 
Nozzle flat into the Contact Tip. The lip on the Nozzle goes into 
the groove on the Contact Tip.

The Contact Tip may also be installed by stretching it over 
the Nozzle.

2

3

Alternate site testing

Alternate site testing means using a part of your body other than 
the fingertips to obtain blood for blood glucose testing. Always 
check with your doctor to see if and when alternate site testing 
is right for you. 

If alternate site testing is right for you, choose a penetration site 
on a soft, fleshy area of your arm, or other recommended site, 
away from bone, and free of visible veins and hair. Obtain a 
sample following Steps 4-12, the same way as shown in 
“Testing from the fingertip”.

Removing the Lancet

Release the Activation Button.

Pull Droplet Genteel away
from your skin

Place Droplet Genteel, gently, but firmly, against 
the fleshy side portion of your middle, ring, or 
pointer finger. Look to be sure there are no 
spaces between the Contact Tip and your skin.

Replace the Nozzle by lining up 
the body’s pins with the Nozzle 
slots and rotate clockwise.

Press the Activation Button and continue to hold firmly 
on the test site. Count to 10 or until you see a blood 
drop form when you look inside the Nozzle window.

9

10

11
12

NOTE:

 Make sure the lancet is inserted 

correctly into the lancing device. It is 

okay for the lancet to turn slightly 

during protective cap removal.

Cleaning Droplet Genteel Lancing Device

1 Surface cleaning: 

Before every use, wipe the outside of the device with a clean 

damp cloth.

IMPORTANT:

 

If the Push Shaft rises past its halfway point – 

while holding the Activation Button, this may be a vacuum leak 

caused by dirt or other contaminants inside Droplet Genteel.

If you experience this vacuum loss symptom – perform (2).

If symptom persists – perform (3). If cleaning doesn't resolve 

vacuum leak, call customer service 1-877-660-1900.

After every use

We recommend using a disinfectant on Droplet Genteel to help 

prevent the transfer of bloodborne viruses or infectious agents. 

We recommend wiping the lancing device, including the inside 

and outside of the Nozzle and Contact Tip, with 70% isopropyl 

alcohol. After wiping, allow Droplet Genteel to air dry completely.

3 Deep clean (inside):

3.1 Remove Piston Assembly:

 Repeat step 2.1

3.2 Flush Vacuum Tube:

 Run fast moving, warm 

water through the Vacuum Tube and out the 

Nozzle while repeatedly pressing the 

Activation Button.

3.3 Flush Piston Assembly:

 While running the 

Piston Assembly under fast-moving, warm 

water repeat 2.2.

3.4 Dry:

 Let Droplet Genteel air dry thoroughly. 

Do not reassemble Droplet Genteel with moisture in the 

Vacuum Tube.

3.5 Reassemble: 

Put Piston Assembly back inside the 

Vacuum Tube and twist base clockwise until it locks into 

place securely. Verify device functions as intended.

2 Quick clean (inside):

2.1 Remove Piston Assembly:

 

De-prime 

Droplet Genteel by pressing Activation 

Button. Twist Piston Assembly base 

counter-clockwise until you can pull it out 

of the Vacuum Tube.

2.2 O-ring Wipe:

 Wipe the O-ring on the 

Piston Assembly with your clean finger to 

remove surface dirt or other contaminants. 

3.2

3.3

2.1

2.2

Distributed by:

HTL-STREFA, Inc.

3005 Chastain Meadows Pkwy, 

Suite 300

Marietta, GA 30066, USA

T: +1-770-528-0410

F: +1-770-528-0411

www.mydroplet.com

[email protected]

Manufactured for:

HTL-STREFA S.A.

ul. Adamówek 7 

95-035 Ozorków, Poland

T: +48 42 270 00 10 

F: +48 42 270 00 20

www.htl-strefa.com

[email protected]

Made in the United States

730/9150/09/2020
20201012_Droplet Genteel Lancing Device_Website IFU

120 Day Satisfaction Guarantee

When you use the Droplet Genteel lancing device with Droplet 30G Personal Lancets and 
in accordance with the instructions for use, we offer a 120 day satisfaction guarantee. 

Please visit www.mydropletgenteel.tips within 30 days of purchase to register your device. 

If you are unsatisfied with your Droplet Genteel lancingdevice for any reason, please call us 
at 1-877-6

6

0-1900 or email us at [email protected] to receive a one-time replacement 

device or refund within 120 days of purchase.

Please do not attempt to return product without prior authorization. Any Droplet Genteel 
lancing device returned without prior authorization will be refused.

If you need troubleshooting assistance or more 
help with your Droplet Genteel lancing device, 

PLEASE DO NOT WAIT

TO CONTACT US.

WE ARE HERE TO HELP!

1-877-660-1900 [email protected]