Droper VET Скачать руководство пользователя страница 11

11/12 

 

KETONES 

 

These aggressive agents may damage the plastic parts and cause the device to 
malfunction.  

 

ALCOHOL-BASED SPRAYS (20% - 40% alcohol content)  

 

These cause the synthetic materials to tarnish and splinter. 

 

IODINE SOLUTIONS 

 

 These may irreversibly stain certain clear plastic parts. 

4.3 

 Re-installing the components 

Once the parts are completely dry, re-insert the two components in the opposite 
manner to the disassembly method, presenting each part opposite its tightening 
clips. The plate must be presented with the infusion bag fastening hooks facing 
the  rear  of  the  device.  Visually  check  the  correct  installation  of  the  different 
elements. 

5.  PERIODIC INSPECTION 

The device is designed to operate more than 6,000 times under normal operating conditions. However, 
the  manufacturer  advises  that  users  perform  annual  maintenance  and  inspections  to  ensure  the 
correct operation of the device, either by sending it to the manufacturer or distributor, or by using the 
method described below

 

:  

Place a 1000 ml bag of NaCl 0.9%, the design of which corresponds to the plate's surface (e.g.

 

: 1000 ml 

Viaflo bag of NaCl 0.9% from Baxter) in the Droper Vet and introduce the spike of a classic infusion line 
without a catheter or needle, and with the valve in the OFF position. Open the infusion line  into a 
recipient placed at the same height as the device and measure the time required to drain the bag. The 
bag must be drained in less than or equal to 10 minutes. 

6.  STORAGE 

The device must be stored in a temperature-controlled, dry place.  

 

Relative humidity

 

: 20% to 90% 

 

Storage temperature

 

: - 10°C to + 50°C 

Содержание VET

Страница 1: ...OPERATING INSTRUCTIONS DROPER VET MAHUSACA sprl Lieu dit Maison Bois 1 4877 OLNE Belgique info droper be ...

Страница 2: ...of contents 1 Description of the Droper Vet 3 2 Using the Droper Vet 4 3 Operating issues 9 4 Cleaning Disinfection 10 5 Periodic inspection 11 6 Storage 11 7 Warranty conditions 12 8 Life 12 9 Useful contacts 12 ...

Страница 3: ...3 12 1 DESCRIPTION OF THE DROPER ...

Страница 4: ...n devices working under pressure like the pressure cuffs Types of infusion lines Two types of lines can be accepted 1 Classic infusion lines similar to these used with the pressure cuffs These lines have normally a drip chamber directly connected to the spike see attached drawing With this accessory there are two possible scenarios a Droper used residentially with flow rate monitored through the d...

Страница 5: ...this accessory there is one possible scenario a Droper used residentially or ambulatory with flow regulation through the drip chamber Possible with all flexible crystalloid bag sizes up to 1000 ml Classical air line purge the Droper can be used vertically hanged or horizontally laid down with its drip chamber directed downward see aside drawing For safety reason any infused animal transfer with a ...

Страница 6: ...the cover then open the cover 2 2 Inserting the infusion line Check that the choke or flow regulator is set to the OFF position Insert the spike into the infusion bag 2 3 Placing the infusion bag connected with the infusion line and closing the device 1 Insertion on the plate With the cover fully open position the bag to be administered on the plate and attach it to one of the fastening hooks on t...

Страница 7: ...ged side down in order to correctly position the cover This operation pressurizes the bag 3 Positioning On the standing horse the Droper vet will be hanged in the upper part of a surcingle see under drawings by the mean of a hook not supplied that will go through the hanging ring installed on the rear side of the Droper Vet in order to benefit from a stable positioning On lying animal ideally the ...

Страница 8: ...he inverted pyramid drawn on the front surface of the mobile plate of the equipment when the final peak will be visible coupled with the absence of flow of infusion solution in the drip chamber if necessary by observance of blood reflux in the tubing with the device being located below the level on which the patient is lying Once administration is complete open the Droper Vet using the same method...

Страница 9: ...ot stopped working Remove the bag of fluid and continue administration by gravity or manual pressure on the bag or change device Send the Droper Vet to the maintenance department for repair Check whether the clamp roller clamp or flow regulator is set to the OFF position Reposition the clamp roller clamp or flow regulator to the desired position Check whether the infusion transfusion line is being...

Страница 10: ...ssembly and cleaning the different parts removed according to the procedure described above in point 4 1 Cleaning can be disinfected by soaking scrubbing or spraying appropriate products The use of disinfectants applied using SPRAYS is ideal for routine disinfection operations and must take place in compliance with the disinfectants manufacturer s recommendations and at a distance of 30 cm from th...

Страница 11: ...o operate more than 6 000 times under normal operating conditions However the manufacturer advises that users perform annual maintenance and inspections to ensure the correct operation of the device either by sending it to the manufacturer or distributor or by using the method described below Place a 1000 ml bag of NaCl 0 9 the design of which corresponds to the plate s surface e g 1000 ml Viaflo ...

Страница 12: ...d or repaired by persons not authorized by us The serial number of the device must not have been modified or erased For item returns and repairs please contact the After Sales Service 8 LIFE The Droper Vet was designed to operate for a period of 2 years calculated based on a daily usage rate of 8 infusions per day from its commissioning date and for a period of 10 years from its date of manufactur...

Отзывы: