background image

Port Washington, NY 11050

V1.1

PDF created with FinePrint pdfFactory trial version 

http://www.pdffactory.com

Содержание AGF-5X

Страница 1: ...Port Washington NY 11050 V1 1 PDF created with FinePrint pdfFactory trial version http www pdffactory com ...

Страница 2: ...es Lead Wires 10 12 Lead Wire Maintenance 10 13 Electrode Options 10 14 Electrode Placement 11 15 Tips For Skin Care 11 16 Application of Re usable Self Adhesive Electrodes 12 17 Adjusting the Controls 13 18 Battery Information 17 19 Maintenance Transportation and Storage of TENS Device 18 20 Safety Control 18 21 Malfunctions 19 22 Conformity to Safety Standards 19 23 Warranty 20 PDF created with ...

Страница 3: ...d flammable mixture Chapter 1 INTRODUCTION EXPLANATION OF PAIN Pain is a warning system and the body s method of telling us that something is wrong Pain is important without it abnormal conditions may go undetected causing damage or injury to vital parts of our bodies Even though pain is a necessary warning signal of trauma or malfunction in the body nature may have gone too far in its design Asid...

Страница 4: ...t effective for pain of central origin This includes headache 4 TENS devices should be used only under the continued supervision of a physician 5 TENS devices have no curative value 6 TENS is a symptomatic treatment and as such suppresses the sensation of pain which would otherwise serve as a protective mechanism 7 Electronic monitoring equipment such as ECG monitors and ECG alarms may not operate...

Страница 5: ...CRIPTION 01 Channel Dual isolated between channels 02 Pulse Ampulitude Adjustable 0 80 mA peak into 500 ohm load each channel 03 Wave Form Asymmetrical Bi Phasic Square Pulse 04 Voltage 0 to 34V Load 500 ohm 05 Power source One 9 Volt Battery Alkaline or nickel cadmium rechargable 06 Size 100 H x 60 W x 21 T mm 07 Weight 120 grams with battery 08 Pulse Rate Adjustable from 1 Hz to 150 Hz 09 Pulse ...

Страница 6: ...e intensity is increased during a low intensity period of the cycle the patient may turn up the control very slowly so that they may feel the intensity any higher TIME DURATION The onset of pain relief should occur shortly after the intensity setting has been determined However in some cases pain relief may take as long as 30 minutes to achieve TENS units are typically operated for long periods of...

Страница 7: ...ethod allowing the current flow on either side of the painful area generally through the nerve branches that feed into the pain site Chapter 15 TIPS FOR SKIN CARE To avoidskin irritation especiallyifyouhave sensitiveskin followthese suggestions 1 Wash the area of skin where you will be placing the electrodes using mild soap and water before applying electrodes and after taking them off Be sure to ...

Страница 8: ...ntrol counter clockwise to the required setting or off position respectively 2 Excess hair may be clipped with scissors do not shave stimulation area 3 Wipe the area with the skin preparation your clinician has recommended Let this dry Apply electrodes as directed 4 Many skin problems arise from the pulling stress from adhesive patches that are excessively stretched across the skin during applicat...

Страница 9: ...me each electrical signal is applied through the skin which controls the strength and sensation of the stimulation If no instructions regarding the pulse width are given in therapy set the control to the suggested 70 120 uS setting 3 Lead Connector Connection of the electrodes is made with two lead connector The device must be switched off before connecting the cables Both intensity controls must ...

Страница 10: ... Always store batteries in their charged condition After a battery has been discharged recharge it as soon as possible If the battery is stored more than 60 days it may need to be recharged c Do not short the terminals of the battery This will cause the battery to get hot and can cause permanent damage Avoid storing the batteries in your pocket or purse where the terminals may accidentally come in...

Страница 11: ...s may leak or burst 2 Do not incinerate the rechargeable battery as it may explode Chapter 19 MAINTENANCE TRANSPORTATION AND STORAGE OF TENS DEVICE 1 Non flammable cleaning solution is suitable for cleaning the device Note Do not smoke or work with open lights for example candles etc when working with flammable liquids 2 Stains and spots can be removed with a cleaning agent 3 Do not submerge the d...

Страница 12: ...ting thereto The warranty does not apply to damage resulting from failure to follow the operating instructions accidents abuse alteration or disassembly by unautho rized personnel Manufacturer Drive Medical Design Manufacturing 99 Seaview Boulevard Port Washington NY 11050 Phone 516 998 4600 Printed in July 2006 PDF created with FinePrint pdfFactory trial version http www pdffactory com ...

Страница 13: ...PDF created with FinePrint pdfFactory trial version http www pdffactory com ...

Отзывы: