25
DL-00096 Rev. 8
Rated maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
150 kHz to 80 MHz
d = 1.2
√
P
0.12
0.38
1.2
3.8
12
Recommended separation distances between portable and mobile RF
communications equipment and the Portable Oxygen Concentrator
The Portable Oxygen Concentrator is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The monitor user can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the monitor as recommended
below, according to the maximum output power of the communications equipment.
Separation distance (m) according to frequency of transmitter
80 MHz to 800 MHz
d = 1.2
√
P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3
√
P
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by the absorption and reflection from structures, objects, and people.
Disclaimer
The information in this document has been carefully examined and is believed
to be reliable. Furthermore, Oxus reserves the right to make changes to any
products herein to improve readability, function, or design. Oxus does not
assume any liability arising out of the application or use of any product or cir
-
cuit described herein; neither does it cover any license under its patent rights
nor the rights of others.