(5) Total Meter
Press
"MENU"
button to select
[TM XXXX.X hr]
menu, the total compliance meter records the
total number of hours that the device has been active. The meter should only be re-set by the
provider, a respiratory therapist or by a physician.
(6) Compliance Meter
Press
"MENU"
button to select
[CM XXXX.X hr]
menu, the compliance meter records the total
therapy hours for the device. The compliance meter should be re-set only by the provider, a
respiratory therapist or by a physician.
(7) Alarm
Press
"MENU"
button to select
[Alarm on/off]
menu, press
"UP"
or
"DOWN"
button to set the
alarm on or off. When alarm is turned on, the audible alarm will activate with warning messages
showed on the LCD display. Set alarm off for mute the audible alarm.
(8) Clock Alarm
Press
"MENU"
button to select
[Clock Alarm on/off]
menu, press
"UP"
or
"DOWN"
button to
set the clock alarm on or off. When clock alarm is set on, the display will show the time on the
left side. Press
"UP"
or
"DOWN"
button to set the time to wake you up. Once the clock alarm is
activated, press the start/standby button to mute the audible alarm.
(9) Clock
Press
"MENU"
button to select
[Clock XX:XX]
menu, press
"UP"
or
"DOWN"
button to set the
current time.
(10) Turn off the Device
Turn off the device by pressing the ON/OFF switch on the left side of the device
NOTE:
Once the setting is confirmed, press “MENU” button. Otherwise, the device will
automatically go back to standby without saving the modification if no action
is taken in 5 seconds.
(11) Event Indication
While the device is on standby mode, press
UP"
and
"DOWN"
button at the same time to see the
latest one data of Apnea Index (AI), Hypopnea Index (HI) and Snoring Index (SI) on the display.
Press
"MENU"
button to show each index in sequence. To go back to standby mode, press
"START/STANDBY"
button.
NOTE: Once the device is re-started, all the indexes will be re-calculated. The whole
data of these indexes are only seen by
respiratory therapists or
physicians by
using Easy Compliance 2.1 playback software.
11