46
Polymax 1, 1/120 °C, 3 and 5
14. Exchange of fuses
Disconnect the unit from mains sup-
ply. The fuse carrier is situated at the
rear of the unit, integrated in the mains
supply.
In order to control or exchange the
fuses push the push catch of the car-
rier, keep it pressed and remove the
carrier from the unit.
Pull the defective fuse out of the clamp
carrier applying slight traction and
replace it by a new, equivalent one.
Important: Only use the fuse
strength indicated on the rear.
After exchanging the fuse insert the
fuse carrier into the housing applying
slight pressure until the fixation latches
noticeably.