3. Ensure that the antenna is located at 0 degrees by adjusting the knob as shown in Figure 7. Click
the
Play
button and a pop-up window will appear. If the antenna is already located at 0 degrees,
click
Ok
to proceed with the measurement. The ME1310 Rx unit will first rotate to 90 degrees,
and the AUT radiation pattern measurement at 0 degrees will then begin.
Figure 7
– Initial angle of AUT
4. Once the measurement is completed, rotate the antenna to 10 degrees by adjusting the knob as
shown in Figure 8.
Figure 8
– AUT Angle rotated to 10 degrees
5. Click
Next
(in the pop-up window shown in Figure 9) and the ME1310 Rx unit will first rotate to
90 degrees, and the AUT radiation pattern measurement at 10 degrees will then begin.
Rotate to
10º