ERROR DISPLAYS AND TROUBLESHOOTING

Conductivity

Display

ATC

Display

y

DISPLAY

Mode

Possible cause(s)

[Action(s)]

"OVER"

"OVER"

Measure

Temperature >99.5

°C range.

[Bring solution to a

lower temperature.]

[Replace

electrode &

ATC assembly .]

"OVER"

"udr"

Measure

Temperature

<0.0°C range.

[Bring solution to a

higher

temperature.]

[Replace

electrode &

ATC assembly .]

"OVER"

60.0 ~ 99.5°C

Measure

Temperature >60.0

°C,

over

the

temperature

compensation

range.

[Bring solution to a

lower temperature.]

"OVER"

0.0 ~ 60.0°C

Measure

The

conductivity

value of the test

solution is beyond

200uS.
[Sample cannot be

tested or replace

electrode & ATC

assembly.]

"ERR"

/

Calibration

a.

Temperature

exceed

0

~

60.0°C

Temperature

compensation

b. Correction of

slope

beyond

50%.

[Bring solution to a

lower temperature.]

[Use

a

new

standard solution.]

[Replace electrode

& ATC assembly.]

Содержание Jenco Vision Plus EC331

Страница 1: ...Operation Manual MODEL EC331 Microcomputer Based Conductivity Resistivity Temperature Pocket Meter EC331...

Страница 2: ...ypad Operations 3 BEFORE YOUR FIRST USE 4 A Insert Batteries 4 B Soak the Electrode 4 C Preparing Standard Solutions 4 D Setup and Calibrate the Electrode and Meter 5 USING VISIONPLUS EC331 5 A Power...

Страница 3: ...materials should be saved until satisfactory operation is confirmed VISIONPLUS EC 331 OVERVIEW A Meter Description 1 Battery cover 2 LCD screen 3 Keypad 4 Electrode collar 5 Electrode ATC assembly Ele...

Страница 4: ...ments 2 Calibration Mode Calibration Mode is used to perform one point calibration 3 Hold Mode Hold Mode is used to display the locked reading for increased ease of use B Keypad Operations Key Operati...

Страница 5: ...is the 147 0uS standard solution the user can prepare to calibrate the probe of the model EC331 Standard solution of 147 0uS at 25 Accurately measure 100mL of the 1413uS conductivity standard solution...

Страница 6: ...ly in distilled water and immerse them in the standard solution of 147uS The temperature displayed is the solution temperature 2 Press Cal key to initiate calibration The CAL icon will appear when the...

Страница 7: ...led water for 10 minutes and recalibrate the EC331 following the instructions detailed in section USING VISIONPLUS EC331 REPLACE THE OLD BATTERIES Replace the battery when the blinking low battery ind...

Страница 8: ...ure Replace electrode ATC assembly OVER 60 0 99 5 C Measure Temperature 60 0 C over the temperature compensation range Bring solution to a lower temperature OVER 0 0 60 0 C Measure The conductivity va...

Страница 9: ...mperature Range Resolution Accuracy 0 0 to 99 5 C 32 to 212 F 0 5 C 1 F 0 5 C 1 F Conductivity Temperature compensation Temperature Coefficient Reference temperature AUTO 0 0 C to 60 0 C 1 91 C 25 C T...

Страница 10: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Страница 11: ...Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Emai...

Отзывы: