- 9 -
Put the samples in the light path and press
Basic Mode
Quantitative
Setup
08:08
08/26
Basic Mode
0.000 Abs
500.0 nm
ESC
ENTER
to measure.
2.9 Print the test result
In the Measurement Interface, press
Basic Mode
Quantitative
Setup
08:08
08/26
Basic Mode
0.000 Abs
500.0 nm
ESC
ENTER
to print the result..
3. Preparation before test
3.1 Switch on the instrument and begin the system self-test(Diagnostic
Program)
Make sure to remove all the blocks in the light path; Close the compartment lid
and switch on the power; Then the system will begin self-test.
Note: During the Diagnostic Program runs, please DON’T open the lid of
the compartment!
3.2 Pre-warm
When the diagnostic program finished, it goes into pre-warm condition. 20
minutes is required before measure.
3.3 Check the Cuvettes
The cuvette should be clear and there
’
s no solution remains on the transparent
surface. If the wavelength is below 340nm,
Silicon Cuvette
is a must.
4. Measure
4.1 Basic Mode
Step 1, Go into Basic Mode interface
In the main menu, use Arrow Key to choose “Baisc Mode”
,
press
Basic Mode
Quantitative
Setup
08:08
08/26
Basic Mode
0.000 Abs
500.0 nm
ESC
ENTER
to go into corresponding setting interface.
(
Fig. 4-2
)
;
Basic Mode
Quantitative
Setup
08:08
08/26
Basic Mode
0.000 Abs
500.0 nm
ESC
ENTER
Fig. 4-2
Step 2, Set Wavelength