2-3
CH 2/ Voltage Measurement Cable and Current Probe Connections
Safety
precautions
The following safety precautions must be followed whenever any type of voltage or
current connection is being made to the PX5.
• Wear proper Personal Protective Equipment, including safety glasses and insulated
gloves when making connections to power circuits.
• Hands, shoes and floor must be dry when making any connection to a power line.
• Before each use, inspect all cables for breaks or cracks in the insulation. Replace
immediately if defective.
• Press the PX5 On/Off power button to Off.
• Before connecting to electric circuits to be monitored, open their related circuit
breakers or disconnects. DO NOT install any connection of the PX5 to live power
lines.
• Connections must be made to the PX5 first, then connect to the circuit to be
monitored.
Continued on next page
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание PowerXplorer PX5
Страница 68: ...3 26 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 84: ...4 16 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 166: ...7 22 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 212: ...A 10 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 226: ...B 14 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 232: ...C 6 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 234: ...D 2 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...
Страница 258: ...F 4 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...