5-26
Turning Monitoring On/Off, continued
Start/Stop
monitoring
(continued)
Continued on next page
Action...
Result...
STEP 2: While the instrument is
actively monitoring data, the
message MONITORING
STATUS: ON appears on the top
section of the Home screen (see
NOTE).
Users cannot change the time and
date nor perform data card
operations while the instrument is
actively monitoring data.
However, the rest of the function
keys, including date style and
clock style, are operable even
while monitoring status is ON.
• To end monitoring, press
Stop
.
Proceed to Step 3.
EP122
NOTE: The following status messages are displayed
after monitoring is turned on.
MONITORING STATUS: ON (monitoring will continue
until Stop is pressed or when specified end time is reached;
START TIME and date are displayed while monitoring is
on)
CIRCUIT TYPE: xxxxxx (default or user specified)
MEMORY CARD: x MB (remaining space in the data card
in MB units)
FILENAME: User specified or Energy Platform Site
xx..ddb (where xx increments every time data is saved using
the same file name)
EVENTS SAVED: xx (counter on the number of events
saved to the data card)
JOURNALS LOGGED: xx (counter on the number of timed
readings recorded)
STEP 3: Stop Monitoring Menu
confirms whether users want to end
monitoring, cancel monitoring, or
view present setups.
• To turn monitoring off, press
Stop
Now!
This will save any pending
data and close the file. Proceed to
Step 4 on page 5-27.
• To cancel monitoring, press
Abort
. All data collected will be
lost when monitoring is aborted.
• To view setups, press
View
present setup
. This will not save
any monitoring parameters.
Proceed to Step 5 on page 5-27.
• To continue monitoring, press
Exit
.
MARK123
Shop for Power Metering products online at:
1.800.561.8187
www.
PowerMeterStore
.ca
Содержание Energy Platform
Страница 13: ...xiii Dranetz Energy PlatformTM Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 14: ...xiv Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 132: ...6 14 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 142: ...7 10 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 146: ...8 4 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 160: ...A 8 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 174: ...B 14 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 180: ...C 6 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 182: ...D 2 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 202: ...F 2 This page intentionally left blank Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...
Страница 223: ...EP124 LEVEL 400 Events EP501 G 19 Shop for Power Metering products online at 1 800 561 8187 www PowerMeterStore ca...