Ultrasonic Flaw Detector DFD30
11
Fine adjustment: +1 or -1mm
Under RANGE, press down key. Right now, there are two tri-angle appear. Under this condition, press + or – key to
realize fine adjustment.
5.1.2 db-STEP
Press left or right key to enter into BASE SETTING, then press up or down key to choose db-STEP.
Press + or – key to change from 01db, 02db, 06db and 12db.
5.1.3 MVTEL
Users are allowed to set the transmission velocity of ultrasonic wave in workpiece. There are two velocities stored in
main unit. They are:
3,240m/s
0.127 in /µs
Sound velocity of transverse wave in steel
5,920m/s
0.233 in /µs
Sound velocity of longitudinal wave in steel
Usera are also allowed to carry our fine adjustment.
Under MVTEL, press down key. Right now, there are two tri-angle appear. Under this condition, press + or – key to
realize fine adjustment.
Fine adjustment: Step is 1m/s 0.0001in/µs
5.1.4 D-DELAY
Users could set the pulse shift during detection, via. D delay. By which, users are allowed to adjust the starting
position for waveform, as well as adjusting the zero point of pulse, so as to make sure that it is at the surface or a
starting face inside the workpiece. If the pulse has to be started from the surface of workpiece, D delay must be set to
0.
Range: 0-512mm
Step:
1mm
There are two ways of adjustment, Rough and fine.
Rough adjustment
Under D-DISPLAY menu, press down key. Right now, there are two tri-angle appear. Under this condition, press + or
– key to realize rough adjustment.
Fine adjustment
Under D-DISPLAY menu, press + or – key directly to realize fine adjustment.
5.1.5 P-ZERO
Range: 0-9.99µs
Step:
0.1µs
There are two ways of adjustment, Rough and fine.
Rough adjustment
Under D-DISPLAY menu, press down key. Right now, there are two tri-angle appear. Under this condition, press + or
– key to realize rough adjustment.