Annexe
73
en
10
Annexe
Optimum test sequence using the Dräger RZ 7000
Residual Pressure Warning
Close cylinder
Alarm at approx. 55 bar
Battery Charge
Remove PSS BG 4 RP from RZ 7000
Deactivate Bodyguard
Insert protective cap
High Pressure Leak Test
(optional)
Open cylinder
≥ 165 bar
Close Open
Constant Metering
Valve & Bypass
Insert protective cap on relief valve
Shortly press red button
flow between 1,5 and 1,9 L/min
Minimum Valve
Remove protective cap
flow noise between
0,1 and 4,5 mbar
Drain Valve
Insert protective cap on relief valve
Pump to fill system
opens between 10 and 25 mbar
Leak Test
Reduce pressure to 7 + 0,5 mbar
Press OK button
pressure difference ≤ 1 mbar/min
Relief Valve
Remove protective cap from
relief valve
Pump to activate
opens between 7 and 9 mbar
Low Pressure Warning
Pump slowly
Alarm before 1,4 mbar
Inhalation Valve
Close exhalation hose
≥ 10 mbar must be reached
Exhalation Valve
Close inhalation hose
≤ -10 mbar must be reached
Battery full
Test Sequence
positive pressure
venting
leak test
negative pressure
L/min
flow
for devices with approval according to Gost R 53259-2009
Содержание PSS BG 4 RP
Страница 1: ...Dr ger PSS BG 4 RP i ru 4 en Instructions for Use 40...
Страница 2: ......
Страница 3: ...9 2 1 3 4 5 6 7 8 10 11 12 14 16 17 18 19 20 21 13 15 1 2 3 4 5 6 7 1 2 3 4 5 5 7 8 33321851 eps B A 6...
Страница 9: ...9 ru 2 7 2 7 1 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger i i...
Страница 12: ...12 ru Dr ger FPS 7000 RP Panorama Nova RP 1 2 2 7 4 PCas APECas 1 2 i i...
Страница 20: ...20 ru 4 1 6 5 6 X X1 X2 X X 3 X CO2 X X4 CO2 X X4 X5 1 2 3 Dr ger 4 CO2 4 5 27 5...
Страница 28: ...28 ru 4 5 1 1 3 2 3 3 3 1 15 4 4 5 2 1 2 3 4 1 4 Bodyguard 1 2 3 4 5 6 7 OK 34921851 eps L min 6 5 3 4 7 1 2...
Страница 35: ...35 ru 5 6 CO2 CO2 7 Dr ger CO2 Dr ger CO2 1 CO2 2 CO2 Dr ger Dr ger CO2 CO2...
Страница 38: ...38 ru PC PCas TR...
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......