Operational Characteristics
12
Make sure that the calibration adapter is correctly seated. To this end, manually
tighten the calibration adapter to a point where the sealing line leaves a permanent
mark on the splash guard.
Nitrogen, synthetic air or fresh air (hydrocarbon content <50 ppm) can be used for
zero point calibration.
Commercially available calibration gas can be used to calibrate the respective gas
category (methane, propane, ethylene). The highest accuracy is achieved using
test gas concentrations of 40 to 70 percent of the measurement span.
The infrared gas transmitter Dräger PIR 3000 can also be used to measure other
gases than mentioned above. For detailed information, refer to "Substitute Gas
Calibration" on page 16.
Select the measured gas in the corresponding table and determine the
corresponding gas category.
Set the gas transmitter to the determined gas category.
Where possible, calibration gas should match with the measured gas for span
calibration. In exceptional cases, span calibration can be carried out using a
suitable substitute gas and the associated calibration factor. The suitable substitute
gas as well as the associated calibration factor is shown in the table "Substitute
Gas Calibration" on page 16.
Select the substitute gas (gas category) and the calibration factor in table
"Substitute Gas Calibration" on page 16.
Multiply the concentration of the substitute gas by the calibration factor to get the
gas concentration to be set.
Example:
Measured gas:
n-octane
Gas category:
Propane (see table "Substitute Gas Calibration",
page 16)
Calibration factor:
1.8 (in table "Substitute Gas Calibration", page 16)
Span gas concentration:
40 %LEL propane (bottle concentration)
Setting:
40 %LEL x 1.8 (calibration factor) = 72 %LEL
CAUTION
Never inhale test gas. Danger to health!
Observe the safety information in the corresponding safety data sheets. Ensure
that gases are vented or otherwise guided outside the building.
Содержание PIR 3000
Страница 1: ...Dräger PIR 3000 en Infrared Gas Transmitter Instructions for Use 00123812_02 eps 0 S ...
Страница 25: ...25 Technical Data Dimensions 01123812_01_en eps 0 S aprox 170 mm ø 37 mm M25 x 1 5 or 3 4 NPT ...
Страница 28: ...ATEX Approval 28 ATEX Approval ...
Страница 29: ...29 ATEX Approval ...
Страница 30: ...ATEX Approval 30 ...
Страница 31: ...31 ATEX Approval ...
Страница 32: ...ATEX Approval 32 ...
Страница 33: ...33 ATEX Approval ...
Страница 34: ...ATEX Approval 34 ...
Страница 35: ...35 ATEX Approval ...
Страница 36: ...ATEX Approval 36 ...
Страница 37: ...37 ATEX Approval ...
Страница 38: ...ATEX Approval 38 ...
Страница 39: ...39 ATEX Approval ...
Страница 40: ...ATEX Approval 40 ...
Страница 41: ...41 ATEX Approval ...
Страница 42: ...ATEX Approval 42 ...
Страница 43: ...43 ATEX Approval ...
Страница 44: ...ATEX Approval 44 ...
Страница 45: ...45 ATEX Approval ...
Страница 46: ...ATEX Approval 46 ...
Страница 47: ...47 ATEX Approval ...
Страница 48: ...IECEx Approval 48 IECEx Approval ...
Страница 49: ...49 IECEx Approval ...
Страница 50: ...IECEx Approval 50 ...
Страница 51: ...51 IECEx Approval ...
Страница 52: ...IECEx Approval 52 ...
Страница 53: ...53 IECEx Approval ...
Страница 54: ...UL Approval 54 UL Approval ...
Страница 55: ...55 UL Approval ...
Страница 56: ...UL Approval 56 ...
Страница 57: ...57 CSA Approval CSA Approval ...
Страница 58: ...CSA Approval 58 ...
Страница 59: ...59 CSA Approval ...
Страница 60: ...Declaration of Conformity 60 Declaration of Conformity ...
Страница 61: ......
Страница 62: ......
Страница 63: ......