Dräger PEX 3000
HQ
_
Operation
– Press the keys (
▲
+
▼
) to select the displayed sensor current as the newly
set sensor current. The value is saved and this function is exited. The
display shows CU once again.
– Press the
▲
key three times to return to measuring mode.
dP – turning the decimal point on
This function is used to select the decimal point in the display if the
DrägerSensor LC M (type XTR 0010 or XTR 0011) or the remote transmitter
type XTR 0090 or XTR 0091 with sensing head SE Ex LC ... DQ is used with a
measuring range of 0 to 9.9 %LEL. The decimal point is only displayed for gas
concentration displays in %LEL.
– Navigate using the
▲
or
▼
keys until the display shows dP ("decimal Point")
and confirm by pressing the keys (
▲
+
▼
). The decimal point is now
displayed if already selected. The flashing point on the bottom right is
switched off.
– Use the
▲
or
▼
key to turn the decimal point on or off.
– Press the keys (
▲
+
▼
) to save the current status and exit the function. The
display shows dP. The activated decimal point is not displayed, rather the
point at the bottom right flashes once again.
– Press the
▲
key four three times to return to measuring mode.
So – displaying the software version
To document necessary software modifications, this function can be used to
display the software version implemented in the transmitter.
– Navigate using the
▲
or
▼
key until the display shows So ("Software") and
confirm by pressing the keys (
▲
+
▼
).
– The display shows a figure between “0.1“ and “9.9“, indicating the
current software version of the transmitter.
– Press the (
▲
+
▼
) keys to exit the function. The display shows So once
again.
– Press the
▲
key five three times to return to measuring mode.
4.4
Commissioning the transmitter
– Open the transmitter’s housing cover.
WARNING
Dust explosion protection is not ensured after having opened the transmitter’s
housing. The Ex area has to be temporarily de-classified, if necessary!
4.4.1
Turning on the supply voltage
– After turning on the supply voltage, all 14 LED segments including both
points are activated for approx. 5 seconds to make sure that they are not
faulty ("Lamp test"). Internal test routines run in the transmitter during this
period, the transmitter’s output signal is 1 mA.
– After this, the display changes, with approx. 1 Hz, between two horizontal
lines and the currently measured gas concentration. During this time the
transmitter’s output is 3.4 mA.
– After one minute the transmitter is ready for operation. If the zero-point is
not undercut by more than 5 %LEL or 1 %LEL, it displays the current
measured value, and the output signal corresponds to the measured value.
– Type XTR 0090 or XTR 0091 with sensing head SE Ex LC M ...DQ: Adjust
the sensor current to 276 mA with the "CU – setting the sensor current"
function (see page 15).
– Allow the transmitter to warm-up for at least 10 minutes (sensor warm-up
time) before calibration.
4.4.2
Calibrating the transmitter
For this chapter:
41507
41508
Содержание PEX 3000
Страница 1: ...Dr ger PEX 3000 de en fr es...
Страница 2: ...de 3 en 27 fr 50 es 74 Dr ger PEX 3000...
Страница 26: ..._ Dr ger PEX 3000 GH _ Konformit tserkl rung 10 Konformit tserkl rung 43496...
Страница 49: ...Declaration of Conformity _ HQ _ Dr ger PEX 3000 10 Declaration of Conformity 43496...
Страница 73: ...D claration de conformit _ IU _ Dr ger PEX 3000 10 D claration de conformit 43496...
Страница 97: ...Declaraci n de conformidad _ HV _ Dr ger PEX 3000 10 Declaraci n de conformidad 43496...
Страница 98: ...HV _ Declaraci n de conformidad _ Dr ger PEX 3000 Esta p gina ha sido dejada en blanco a prop sito...
Страница 99: ...Declaraci n de conformidad _ HV _ Dr ger PEX 3000 Esta p gina ha sido dejada en blanco a prop sito...