Page 36 / 38
9
No sample gas line is available
Use infusion line or CO
2
measurement line with LuerLock
Gas measurement might be affected, especialy when applying higher respiratory rates and
lower tidal volumes
Alarms like
„sample line blocked“ or „sample line disconnected?” might appear
10 No breathing system filter is available
General information on the filter setup and necessary reprocessing measures are in the customer letter
“SARS-CoV-2 and handling of Dräger Anesthesia Workstations“,
https://www.draeger.com/en-
us_us/Home/novel-coronavirus-outbreak
Overview of Dräger filter / HME types and abbreviations for the decision tree:
MF/HME
Mechanical filter with HME = Dräger TwinStar HEPA
MF
Mechanical filter = Dräger SafeStar
EF/HME
Electrostatic filter with HME = Dräger TwinStar
EF
Electrostatic filter = Dräger CareStar
HME
HME only (no filter) = Dräger HumidStar