34
Instructions for use
|
Oxylog VE300 SW 1.n
Assembly and preparation
Assembly
1. Connect a pressure reducer to the oxygen cylinder or use an oxygen cylinder
with a built-in pressure reducer.
2. Open the device closure and open the device.
3. Place the cylinder in the well of the base plate and position the oxygen cylinder
so that the pressure reducer is on the side of the bar and the cylinder body is
resting on all 4 rubber feet.
4. Adjust the hinge and closure length once to the diameter of the oxygen cylinder:
Remove the upper screw of the hinge with a hexagon key, remove the protective
plate and loosen the 2 screws beneath the plate.
Set the catch (A) so that the ventilator is resting horizontally on the cylinder, and
then re-tighten the screws. Refit the protective plate.
Loosen the 2 screws of the closure, set the closure to the same scale value as
the hinge, and then re-tighten the screws. Make sure that the closures do not
wiggle easily when closed and adjust the catch to a lower position if necessary.
WARNING
Risk of patient injury
A covering on the oxygen cylinder (e.g., a net) prevents sufficient fixation of the
oxygen cylinder in the ventilator.
►
Make sure that the oxygen cylinder is mounted in the ventilator with no
covering over it.
Standard protection bar
Wide protection bar
33375